Agilent E5250A User’s Guide, Edition 11 2-11
Installation
Installing the E5250A
Figure 2-2 Agilent E5255A Component Locations
To Modify BIAS Input
Each multiplexer block has an independent BIAS input. You can change the BIAS
input internally using the BIASin/BIASout terminals shown in Figure 2-2, or
externally using the BIAS INPUT connectors. For actual examples of modifications,
see “E5255A Configuration Examples”.
To Use Multiplexer Blocks on another E5255A
When a module is installed in a mainframe slot, six module pins are connected to
the mainframe internal bus. So, if multiplexers on different E5255As use the same
module pin, the IV inputs are connected together via the internal bus.