Agilent TechnologiesAgilent E5250ALow Leakage Switch MainframeUser’s Guide
5-6 Agilent E5250A User’s Guide, Edition 11Controlling the E5250ASwitch Control BasicsChannel List and Channel Configuration Some SCPI commands requir
Agilent E5250A User’s Guide, Edition 11 5-7Controlling the E5250ASwitch Control BasicsTable 5-2 Card Number In the channel_list, you can specify multi
5-8 Agilent E5250A User’s Guide, Edition 11Controlling the E5250ASwitch Control BasicsChannel ConfigurationThe E5250A allows the following two channel
Agilent E5250A User’s Guide, Edition 11 5-9Controlling the E5250ASwitch Control BasicsConnection Rule Connection rule is used to specify if an input o
5-10 Agilent E5250A User’s Guide, Edition 11Controlling the E5250ASwitch Control BasicsConnection Sequence NOTE This is only for cards that are set to
Agilent E5250A User’s Guide, Edition 11 5-11Controlling the E5250ASwitch Control BasicsBias Mode NOTE Connection rule cannot be specified for the inpu
5-12 Agilent E5250A User’s Guide, Edition 11Controlling the E5250ASwitch Control BasicsFigure 5-2 Example of Bias Mode Figure 5-2 shows an example. In
Agilent E5250A User’s Guide, Edition 11 5-13Controlling the E5250ASwitch Control BasicsCouple Port NOTE This function is available only for the E5252A
5-14 Agilent E5250A User’s Guide, Edition 11Controlling the E5250ASwitch Control BasicsThe two output ports are determined by the channel_list paramet
Agilent E5250A User’s Guide, Edition 11 5-15Controlling the E5250ASwitch Control BasicsE5252A Channel List This section describes how to define the ch
ContentsAgilent E5250A User’s Guide, Edition 11 Contents - 11. Introduction Agilent E5250A Product Description . . . . . . . . . . . . . . . . . . .
5-16 Agilent E5250A User’s Guide, Edition 11Controlling the E5250ASwitch Control BasicsFigure 5-4 E5252A Channel Number in Normal Config. Mode Input
Agilent E5250A User’s Guide, Edition 11 5-17Controlling the E5250ASwitch Control BasicsAuto Configuration ModeTo use this mode, the E5252A must be ins
5-18 Agilent E5250A User’s Guide, Edition 11Controlling the E5250ASwitch Control BasicsFigure 5-5 E5252A Channel Number in Auto Config. Mode (example
Agilent E5250A User’s Guide, Edition 11 5-19Controlling the E5250ASwitch Control BasicsE5255A Channel List This section describes how to define the ch
5-20 Agilent E5250A User’s Guide, Edition 11Controlling the E5250ASwitch Control BasicsCard Number: One digit: 1 to 4. E5250A slot number in which ca
Agilent E5250A User’s Guide, Edition 11 5-21Controlling the E5250ASwitch Control BasicsAuto Configuration ModeTo use this mode, the E5255A must be ins
5-22 Agilent E5250A User’s Guide, Edition 11Controlling the E5250ASwitch Control BasicsIn the Auto Configuration mode, you can omit any zeros at begin
Agilent E5250A User’s Guide, Edition 11 5-23Controlling the E5250ASwitch Control BasicsTable 5-9 Example IV Input Connection and Channel NumberSlot No
5-24 Agilent E5250A User’s Guide, Edition 11Controlling the E5250ASwitch Control BasicsFigure 5-8 Example for E5255A Channel Number in Auto Config. Mo
Agilent E5250A User’s Guide, Edition 11 5-25Controlling the E5250AUsing Virtual Front Panel UtilityUsing Virtual Front Panel Utility The Virtual Front
Contents - 2 Agilent E5250A User’s Guide, Edition 11ContentsMaintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5-26 Agilent E5250A User’s Guide, Edition 11Controlling the E5250AUsing Virtual Front Panel UtilityFunctions of Virtual Front Panel Utility The Virtua
Agilent E5250A User’s Guide, Edition 11 5-27Controlling the E5250AUsing Virtual Front Panel UtilityRequirements The following are required to execute
5-28 Agilent E5250A User’s Guide, Edition 11Controlling the E5250AUsing Virtual Front Panel UtilityThe following softkeys are displayed on right side
Agilent E5250A User’s Guide, Edition 11 5-29Controlling the E5250AUsing Virtual Front Panel UtilityChanging the VFP Setup Mode Select the SETUP MODE s
5-30 Agilent E5250A User’s Guide, Edition 11Controlling the E5250AUsing Virtual Front Panel UtilityFigure 5-9 Example of SETUP MODE MENU Agilent E5250
Agilent E5250A User’s Guide, Edition 11 5-31Controlling the E5250AUsing Virtual Front Panel UtilityDefining Labels for the E5250A Input Ports You can
5-32 Agilent E5250A User’s Guide, Edition 11Controlling the E5250AUsing Virtual Front Panel UtilityTo define labels for the E5255A bias input portsSel
Agilent E5250A User’s Guide, Edition 11 5-33Controlling the E5250AUsing Virtual Front Panel UtilityMaking Connections and Changing the Setup Data Sele
5-34 Agilent E5250A User’s Guide, Edition 11Controlling the E5250AUsing Virtual Front Panel UtilityFigure 5-11 Example of SETUP MEMORY MENU Agilent E5
Agilent E5250A User’s Guide, Edition 11 5-35Controlling the E5250AUsing Virtual Front Panel UtilityTo delete a setup memorySelect the DELETE SETUP sof
ContentsAgilent E5250A User’s Guide, Edition 11 Contents - 3To Connect the E5252A Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5-36 Agilent E5250A User’s Guide, Edition 11Controlling the E5250AUsing Virtual Front Panel UtilityTo modify the settings in Current Setup MemorySelec
Agilent E5250A User’s Guide, Edition 11 5-37Controlling the E5250AUsing Virtual Front Panel UtilityFigure 5-13 Memory Setup Panel for the E5255A in No
5-38 Agilent E5250A User’s Guide, Edition 11Controlling the E5250AUsing Virtual Front Panel UtilityYou can define labels to use for the input ports of
Agilent E5250A User’s Guide, Edition 11 5-39Controlling the E5250AUsing Virtual Front Panel UtilityTo enter a comment for the setup memorySelect the C
5-40 Agilent E5250A User’s Guide, Edition 11Controlling the E5250AUsing Virtual Front Panel UtilitySoftkey Menu 1:NOTE Moving the Cursor Move the curs
Agilent E5250A User’s Guide, Edition 11 5-41Controlling the E5250AUsing Virtual Front Panel UtilitySoftkey Menu 2:For the following conditions, the fo
5-42 Agilent E5250A User’s Guide, Edition 11Controlling the E5250AUsing Virtual Front Panel UtilitySaving/Loading the Setup Data Select the SAVE/LOAD
6 Programming the E5250A
6-2 Agilent E5250A User’s Guide, Edition 11Programming the E5250AStandard Commands for Programmable Instruments (SCPI) is a universal programming lang
Agilent E5250A User’s Guide, Edition 11 6-3Programming the E5250ASCPI Programming BasicsSCPI Programming Basics This section describes the structure o
Contents - 4 Agilent E5250A User’s Guide, Edition 11ContentsControlling E5250A by Using HP BASIC . . . . . . . . . . . . . . . . . . . . . . . . . .
6-4 Agilent E5250A User’s Guide, Edition 11Programming the E5250ASCPI Programming BasicsSCPI Command Hierarchy The SCPI commands use a hierarchical st
Agilent E5250A User’s Guide, Edition 11 6-5Programming the E5250ASCPI Programming BasicsControlling E5250A by Using HP BASIC 1. Assign I/O path for co
6-6 Agilent E5250A User’s Guide, Edition 11Programming the E5250ACreating a Control ProgramCreating a Control Program This section explains how to cre
Agilent E5250A User’s Guide, Edition 11 6-7Programming the E5250ACreating a Control ProgramTable 6-1 Programming Summary For details about these param
6-8 Agilent E5250A User’s Guide, Edition 11Programming the E5250ACreating a Control ProgramDefining Channel Configuration Mode Following commands are
Agilent E5250A User’s Guide, Edition 11 6-9Programming the E5250ACreating a Control ProgramDefining Connection Sequence Following commands are used to
6-10 Agilent E5250A User’s Guide, Edition 11Programming the E5250ACreating a Control ProgramUsing Bias Mode If you use the Bias Mode, following comman
Agilent E5250A User’s Guide, Edition 11 6-11Programming the E5250ACreating a Control ProgramUsing Couple Port (for E5252A only) If you use Couple Port
6-12 Agilent E5250A User’s Guide, Edition 11Programming the E5250ACreating a Control ProgramControlling Relay Switching Following commands are used to
Agilent E5250A User’s Guide, Edition 11 6-13Programming the E5250AProgramming ExamplesProgramming Examples This section provides examples of control p
ContentsAgilent E5250A User’s Guide, Edition 11 Contents - 5Status Reporting Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6-14 Agilent E5250A User’s Guide, Edition 11Programming the E5250AProgramming ExamplesChannel Creation Example The following example is a typical cont
Agilent E5250A User’s Guide, Edition 11 6-15Programming the E5250AProgramming ExamplesProgram List: 10 ! Example: Creating Connection Channels 20 ! 3
6-16 Agilent E5250A User’s Guide, Edition 11Programming the E5250AProgramming ExamplesBias Mode Example The following example shows how to use Bias Mo
Agilent E5250A User’s Guide, Edition 11 6-17Programming the E5250AProgramming ExamplesProgram List: 10 ! Example: Using Bias Mode 20 ! 30 INTEGER Com
6-18 Agilent E5250A User’s Guide, Edition 11Programming the E5250AProgramming Examples160 ! 170 ! Put Bias Source Control Routine here. 180 ! 190 OUTP
Agilent E5250A User’s Guide, Edition 11 6-19Programming the E5250AProgramming ExamplesCouple Port Example The following example shows how to use the C
6-20 Agilent E5250A User’s Guide, Edition 11Programming the E5250AProgramming ExamplesProgram List: 10 ! Example: Using Couple Port 20 ! 30 INTEGER
Agilent E5250A User’s Guide, Edition 11 6-21Programming the E5250AProgramming Examples150 OUTPUT @Hp5250;":ROUT:COUP:PORT ALL,'3,5'&quo
6-22 Agilent E5250A User’s Guide, Edition 11Programming the E5250AUsing the VFP Data Upload LibraryUsing the VFP Data Upload Library This section intr
Agilent E5250A User’s Guide, Edition 11 6-23Programming the E5250AUsing the VFP Data Upload LibraryVFP Data Upload Library VFP Data Upload Library is
Contents - 6 Agilent E5250A User’s Guide, Edition 11Contents12. SCPI Command Summary
6-24 Agilent E5250A User’s Guide, Edition 11Programming the E5250AUsing the VFP Data Upload LibraryConnect_vfp subprogramMakes relay connections based
Agilent E5250A User’s Guide, Edition 11 6-25Programming the E5250AUsing the VFP Data Upload LibraryProgramming Example The following program is an exa
6-26 Agilent E5250A User’s Guide, Edition 11Programming the E5250AUsing the VFP Data Upload LibraryBefore Executing Your Program Before executing a pr
Agilent E5250A User’s Guide, Edition 11 6-27Programming the E5250AUsing the VFP Data Upload LibraryError Messages If any error related to the VFP Data
6-28 Agilent E5250A User’s Guide, Edition 11Programming the E5250AUsing the Capacitance Compensation RoutineUsing the Capacitance Compensation Routine
Agilent E5250A User’s Guide, Edition 11 6-29Programming the E5250AUsing the Capacitance Compensation RoutineCapacitance Compensation Routine This rout
6-30 Agilent E5250A User’s Guide, Edition 11Programming the E5250AUsing the Capacitance Compensation RoutineRequired Conditions This subprogram requir
Agilent E5250A User’s Guide, Edition 11 6-31Programming the E5250AUsing the Capacitance Compensation Routine1. Measure R, L, C of your cable using the
6-32 Agilent E5250A User’s Guide, Edition 11Programming the E5250AUsing the Capacitance Compensation RoutineProgramming Example The following program
7 Command Reference
1 Introduction
7-2 Agilent E5250A User’s Guide, Edition 11Command ReferenceThis chapter describes the following for Agilent E5250A:• SCPI commands available to contr
Agilent E5250A User’s Guide, Edition 11 7-3Command ReferenceTextual Notation for Subsystem Commands CAPITAL LETTERS Capital letters are the minimally
7-4 Agilent E5250A User’s Guide, Edition 11Command ReferenceCommon Commands for E5250ACommon Commands This section describes common commands and queri
Agilent E5250A User’s Guide, Edition 11 7-5Command Reference*CLS*CLS This command clears the Status Byte Register, the Standard Event Status Register,
7-6 Agilent E5250A User’s Guide, Edition 11Command Reference*ESEThe following table shows the bits of the Standard Event Status Register and binary-we
Agilent E5250A User’s Guide, Edition 11 7-7Command Reference*ESR?*ESR? This query command returns the present contents of the Standard Event Status Re
7-8 Agilent E5250A User’s Guide, Edition 11Command Reference*IDN?*IDN? This query command returns the ID of your E5250A.Syntax *IDN?Query response H
Agilent E5250A User’s Guide, Edition 11 7-9Command Reference*OPC*OPC This command starts to monitor pending operations, and sets/clears the Operation
7-10 Agilent E5250A User’s Guide, Edition 11Command Reference*RST*RST This command performs an instrument reset. Status after *RST is shown below:Chan
Agilent E5250A User’s Guide, Edition 11 7-11Command Reference*SRE*SRE This command sets the Service Request "Enable" Register bits. 1 enable
1-2 Agilent E5250A User’s Guide, Edition 11IntroductionThis chapter gives an overview of Agilent E5250A, E5252A, and E5255A.• “Agilent E5250A Product
7-12 Agilent E5250A User’s Guide, Edition 11Command Reference*SREExample The following four lines enable the same bits (bit 4 and 5):OUTPUT @Hp5250;&
Agilent E5250A User’s Guide, Edition 11 7-13Command Reference*STB?*STB? This query command reads the Status Byte Register (reads Master Summary Status
7-14 Agilent E5250A User’s Guide, Edition 11Command Reference*TST?*TST? This query command executes an internal self-test, then returns the result. Af
Agilent E5250A User’s Guide, Edition 11 7-15Command Reference:DIAGnostic Command Tree:DIAGnostic Subsystem The :DIAGnostic subsystem has commands for
7-16 Agilent E5250A User’s Guide, Edition 11Command Reference:DIAGnostic:TEST:CARD:CLEar:DIAGnostic:TEST:CARD:CLEar This command clears the relay test
Agilent E5250A User’s Guide, Edition 11 7-17Command Reference:DIAGnostic:TEST:CARD:STATe?:DIAGnostic:TEST:CARD:STATe? This command returns the most re
7-18 Agilent E5250A User’s Guide, Edition 11Command Reference:DIAGnostic:TEST:FRAMe[:EXECute]?:DIAGnostic:TEST:FRAMe[:EXECute]? This command executes
Agilent E5250A User’s Guide, Edition 11 7-19Command Reference:DIAGnostic:TEST:FRAMe:STATe?:DIAGnostic:TEST:FRAMe:STATe? This command returns the most
7-20 Agilent E5250A User’s Guide, Edition 11Command Reference:ROUTe Command Tree:ROUTe Subsystem The :ROUTe subsystem has commands for controlling the
Agilent E5250A User’s Guide, Edition 11 7-21Command Reference:ROUTe Command TreeParameter values:card_number: Auto Config Mode: 0 Normal Config Mode:
Agilent E5250A User’s Guide, Edition 11 1-3IntroductionAgilent E5250A Product DescriptionAgilent E5250A Product Description Agilent E5250A Low Leakage
7-22 Agilent E5250A User’s Guide, Edition 11Command Reference[:ROUTe]:BIAS:CHANnel:DISable:CARD[:ROUTe]:BIAS:CHANnel:DISable:CARD This command bias di
Agilent E5250A User’s Guide, Edition 11 7-23Command Reference[:ROUTe]:BIAS:CHANnel:DISable[:LIST]?[:ROUTe]:BIAS:CHANnel:DISable[:LIST]? This query com
7-24 Agilent E5250A User’s Guide, Edition 11Command Reference[:ROUTe]:BIAS:CHANnel:ENABle:CARD[:ROUTe]:BIAS:CHANnel:ENABle:CARD This command bias enab
Agilent E5250A User’s Guide, Edition 11 7-25Command Reference[:ROUTe]:BIAS:CHANnel:ENABle[:LIST][:ROUTe]:BIAS:CHANnel:ENABle[:LIST] This command bias
7-26 Agilent E5250A User’s Guide, Edition 11Command Reference[:ROUTe]:BIAS:CHANnel:ENABle[:LIST]?[:ROUTe]:BIAS:CHANnel:ENABle[:LIST]? This query comma
Agilent E5250A User’s Guide, Edition 11 7-27Command Reference[:ROUTe]:BIAS:PORT[:ROUTe]:BIAS:PORT NOTE This command applies only to the E5252A card.Fo
7-28 Agilent E5250A User’s Guide, Edition 11Command Reference[:ROUTe]:BIAS:PORT?[:ROUTe]:BIAS:PORT? NOTE This command applies only to the E5252A card.
Agilent E5250A User’s Guide, Edition 11 7-29Command Reference[:ROUTe]:BIAS[:STATe][:ROUTe]:BIAS[:STATe] This command controls the Bias Mode for the sp
7-30 Agilent E5250A User’s Guide, Edition 11Command Reference[:ROUTe]:BIAS[:STATe]?[:ROUTe]:BIAS[:STATe]? This query command returns the Bias Mode of
Agilent E5250A User’s Guide, Edition 11 7-31Command Reference[:ROUTe]:CLOSe[:LIST][:ROUTe]:CLOSe[:LIST] This command connects the input ports to the o
Notices© Agilent Technologies, Inc. 1995 - 2011No part of this manual may be reproduced in any form or by any means (including electronic storage and
1-4 Agilent E5250A User’s Guide, Edition 11IntroductionAgilent E5250A Product DescriptionFront Panel Tour The E5250A has 2 hard keys and 4 LED indicat
7-32 Agilent E5250A User’s Guide, Edition 11Command Reference[:ROUTe]:CONNection:RULE[:ROUTe]:CONNection:RULE This command sets the connection rule (F
Agilent E5250A User’s Guide, Edition 11 7-33Command Reference[:ROUTe]:CONNection:SEQuence[:ROUTe]:CONNection:SEQuence This command specifies the conne
7-34 Agilent E5250A User’s Guide, Edition 11Command Reference[:ROUTe]:CONNection:SEQuence?[:ROUTe]:CONNection:SEQuence? This query command returns the
Agilent E5250A User’s Guide, Edition 11 7-35Command Reference[:ROUTe]:COUPle:PORT[:ROUTe]:COUPle:PORT NOTE This command applies only to the E5252A car
7-36 Agilent E5250A User’s Guide, Edition 11Command Reference[:ROUTe]:COUPle:PORT?[:ROUTe]:COUPle:PORT? NOTE This command applies only to the E5252A c
Agilent E5250A User’s Guide, Edition 11 7-37Command Reference[:ROUTe]:COUPle[:STATe][:ROUTe]:COUPle[:STATe] NOTE This command applies only to the E525
7-38 Agilent E5250A User’s Guide, Edition 11Command Reference[:ROUTe]:COUPle[:STATe]?[:ROUTe]:COUPle[:STATe]? NOTE This command applies only to the E5
Agilent E5250A User’s Guide, Edition 11 7-39Command Reference[:ROUTe]:FUNCtion[:ROUTe]:FUNCtion This command sets the channel configuration, which det
7-40 Agilent E5250A User’s Guide, Edition 11Command Reference[:ROUTe]:OPEN:CARD[:ROUTe]:OPEN:CARD This command disconnects all input ports from all ou
Agilent E5250A User’s Guide, Edition 11 7-41Command Reference[:ROUTe]:OPEN[:LIST]?[:ROUTe]:OPEN[:LIST]? This query command returns 0 or 1 for each cha
Agilent E5250A User’s Guide, Edition 11 1-5IntroductionAgilent E5250A Product DescriptionAUX INPUT AUX INPUT connectors are for C-V measurements, pul
7-42 Agilent E5250A User’s Guide, Edition 11Command Reference:SYSTem Command Tree:SYSTem Subsystem The :SYSTem subsystem is a collection of functions
Agilent E5250A User’s Guide, Edition 11 7-43Command Reference:SYSTem:CCONfig?:SYSTem:CCONfig? This query command returns the card configuration inform
7-44 Agilent E5250A User’s Guide, Edition 11Command Reference:SYSTem:CDEScription?:SYSTem:CDEScription? This query command returns a description of th
Agilent E5250A User’s Guide, Edition 11 7-45Command Reference:SYSTem:CPON:SYSTem:CPON This command resets the specified card to the power-on state. If
7-46 Agilent E5250A User’s Guide, Edition 11Command Reference:SYSTem:CTYPe?:SYSTem:CTYPe? This query command returns the ID of the specified card.This
Agilent E5250A User’s Guide, Edition 11 7-47Command Reference:SYSTem:ERRor?:SYSTem:ERRor? This command reads the error from the head of the error queu
7-48 Agilent E5250A User’s Guide, Edition 11Command ReferenceStatus Reporting StructureStatus Reporting Structure This section describes the status re
Agilent E5250A User’s Guide, Edition 11 7-49Command ReferenceStatus Reporting StructureFigure 7-1 Status Reporting Structure of E5250A The above figur
7-50 Agilent E5250A User’s Guide, Edition 11Command ReferenceStatus Byte RegisterStatus Byte Register The E5250A Status Byte Register contains bits (E
Agilent E5250A User’s Guide, Edition 11 7-51Command ReferenceStatus Byte RegisterIn general, use serial polling (not *STB?) inside interrupt service r
1-6 Agilent E5250A User’s Guide, Edition 11IntroductionAgilent E5252A Product DescriptionAgilent E5252A Product Description The E5252A 10×12 Matrix Sw
7-52 Agilent E5250A User’s Guide, Edition 11Command ReferenceService Request Enable RegisterService Request Enable Register The Service Request Enable
Agilent E5250A User’s Guide, Edition 11 7-53Command ReferenceStandard Event Status RegisterStandard Event Status Register The Standard Event Status Re
7-54 Agilent E5250A User’s Guide, Edition 11Command ReferenceStandard Event Status Enable RegisterTable 7-2 Standard Event Status Register of E5250A S
Agilent E5250A User’s Guide, Edition 11 7-55Command ReferenceOutput QueueOutput Queue The Output Queue stores response messages until they are read. I
7-56 Agilent E5250A User’s Guide, Edition 11Command ReferenceOutput Queue
8 VXIplug&play Driver
8-2 Agilent E5250A User’s Guide, Edition 11VXIplug&play DriverThis chapter introduces the VXIplug&play driver available for Agilent E5250A, an
Agilent E5250A User’s Guide, Edition 11 8-3VXIplug&play DriverSystem RequirementsSystem RequirementsThe following system environments are required
8-4 Agilent E5250A User’s Guide, Edition 11VXIplug&play DriverInstalling VXIplug&play DriverInstalling VXIplug&play DriverThe installation
Agilent E5250A User’s Guide, Edition 11 8-5VXIplug&play DriverDriver FunctionsDriver FunctionsTable 8-1 lists all the functions for the E5250A. Yo
Agilent E5250A User’s Guide, Edition 11 1-7IntroductionAgilent E5252A Product DescriptionInput Ports The E5252A has 10 input ports via the E5250A.Inp
8-6 Agilent E5250A User’s Guide, Edition 11VXIplug&play DriverDriver FunctionsRoute Control hpe5250a_closeList Closes the channel list.hpe5250a_op
Agilent E5250A User’s Guide, Edition 11 8-7VXIplug&play DriverDriver Functionshpe5250a_biasChanCardThis function will enable or disable bias on al
8-8 Agilent E5250A User’s Guide, Edition 11VXIplug&play DriverDriver Functionshpe5250a_biasChanList_QThis function will query the instrument for t
Agilent E5250A User’s Guide, Edition 11 8-9VXIplug&play DriverDriver Functionshpe5250a_biasPortThis function will select which input port is the b
8-10 Agilent E5250A User’s Guide, Edition 11VXIplug&play DriverDriver Functionshpe5250a_closeThis function terminates the software connection to t
Agilent E5250A User’s Guide, Edition 11 8-11VXIplug&play DriverDriver Functionshpe5250a_closeListThis function will connect the input ports to the
8-12 Agilent E5250A User’s Guide, Edition 11VXIplug&play DriverDriver Functionshpe5250a_closeList_QThis function will query the instrument for the
Agilent E5250A User’s Guide, Edition 11 8-13VXIplug&play DriverDriver Functionshpe5250a_cmdData_QThis function passes the cmd_str string to the in
8-14 Agilent E5250A User’s Guide, Edition 11VXIplug&play DriverDriver Functionshpe5250a_cmdInt16Arr_QThis function passes the cmd_str string to th
Agilent E5250A User’s Guide, Edition 11 8-15VXIplug&play DriverDriver Functionshpe5250a_cmdInt32Arr_QThis function passes the cmd_str string to th
1-8 Agilent E5250A User’s Guide, Edition 11IntroductionAgilent E5255A Product DescriptionAgilent E5255A Product Description The E5255A 24 (8×3) Channe
8-16 Agilent E5250A User’s Guide, Edition 11VXIplug&play DriverDriver Functionshpe5250a_cmdRealThis function passes the cmd_str string to the inst
Agilent E5250A User’s Guide, Edition 11 8-17VXIplug&play DriverDriver Functionshpe5250a_cmdReal64_QThis function passes the cmd_str string to the
8-18 Agilent E5250A User’s Guide, Edition 11VXIplug&play DriverDriver Functionshpe5250a_compenCThis function compensates capacitance/conductance d
Agilent E5250A User’s Guide, Edition 11 8-19VXIplug&play DriverDriver Functionshpe5250a_connRuleSeqThe function sets connection rule and connectio
8-20 Agilent E5250A User’s Guide, Edition 11VXIplug&play DriverDriver Functionshpe5250a_couplePortThis function sets the couple ports which are us
Agilent E5250A User’s Guide, Edition 11 8-21VXIplug&play DriverDriver Functionshpe5250a_coupleStateThis function controls the couple port mode for
8-22 Agilent E5250A User’s Guide, Edition 11VXIplug&play DriverDriver Functionshpe5250a_error_messageThis function translates the error return val
Agilent E5250A User’s Guide, Edition 11 8-23VXIplug&play DriverDriver Functionshpe5250a_errorQueryDetectThis function enables or disables automati
8-24 Agilent E5250A User’s Guide, Edition 11VXIplug&play DriverDriver Functionshpe5250a_esr_QThis function returns the contents of the ESR registe
Agilent E5250A User’s Guide, Edition 11 8-25VXIplug&play DriverDriver Functionshpe5250a_initThis function initializes the software connection to t
Agilent E5250A User’s Guide, Edition 11 1-9IntroductionAgilent E5255A Product DescriptionResistors Resistors connected between BIAS path and IV path
8-26 Agilent E5250A User’s Guide, Edition 11VXIplug&play DriverDriver Functionshpe5250a_openCardThis function will disconnect all input ports from
Agilent E5250A User’s Guide, Edition 11 8-27VXIplug&play DriverDriver Functionshpe5250a_openList_QThis function will query the instrument for the
8-28 Agilent E5250A User’s Guide, Edition 11VXIplug&play DriverDriver Functionshpe5250a_resetThis function places the instrument in a default stat
Agilent E5250A User’s Guide, Edition 11 8-29VXIplug&play DriverDriver Functionshpe5250a_selectCompenFileThis function specifies capacitance/conduc
8-30 Agilent E5250A User’s Guide, Edition 11VXIplug&play DriverDriver Functionshpe5250a_self_testThis function causes the instrument to perform a
Agilent E5250A User’s Guide, Edition 11 8-31VXIplug&play DriverDriver Functionshpe5250a_testExec_QThis function executes the controller test, the
8-32 Agilent E5250A User’s Guide, Edition 11VXIplug&play DriverDriver Functions
9 Executing Sample Programs
9-2 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsThis chapter explains how to use the following sample programs:• “Vth and Capacita
Agilent E5250A User’s Guide, Edition 11 9-3Executing Sample ProgramsVth and Capacitance Measurement ProgramVth and Capacitance Measurement Program Thi
1-10 Agilent E5250A User’s Guide, Edition 11IntroductionAgilent E5255A Product DescriptionFigure 1-5 E5255A Block Diagram BIAS Input 1OutputResistorIV
9-4 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsVth and Capacitance Measurement ProgramTable 9-2 Vth and Capacitance Measurement F
Agilent E5250A User’s Guide, Edition 11 9-5Executing Sample ProgramsVth and Capacitance Measurement ProgramVth Measurement SetupMeasurement setup of t
9-6 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsVth and Capacitance Measurement ProgramDisplay Setup: Also the DATA VARIAVLES fiel
Agilent E5250A User’s Guide, Edition 11 9-7Executing Sample ProgramsVth and Capacitance Measurement ProgramSetting up the Measurement Environment Befo
9-8 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsVth and Capacitance Measurement ProgramFigure 9-1 Vth and C Measurement Setup and
Agilent E5250A User’s Guide, Edition 11 9-9Executing Sample ProgramsVth and Capacitance Measurement ProgramExecuting the Program Before executing the
9-10 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsVth and Capacitance Measurement ProgramExecute the program as follows:1. LOAD the
Agilent E5250A User’s Guide, Edition 11 9-11Executing Sample ProgramsVth and Capacitance Measurement ProgramModifying the Program This section explain
9-12 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsVth and Capacitance Measurement ProgramTo output the measurement report to printe
Agilent E5250A User’s Guide, Edition 11 9-13Executing Sample ProgramsVth and Capacitance Measurement ProgramTo change the Vth measurement setupYou can
Agilent E5250A User’s Guide, Edition 11 1-11IntroductionAgilent E5255A Product DescriptionTable 1-2 E5255A Input/Output Connectors Figure 1-6 E5255A B
9-14 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsVth and Capacitance Measurement ProgramTo change the capacitance measurement setu
Agilent E5250A User’s Guide, Edition 11 9-15Executing Sample ProgramsVth and Capacitance Measurement ProgramTo use the VFP data upload libraryIf you w
9-16 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsVth and Capacitance Measurement Program2. Remove the exclamation marks from progr
Agilent E5250A User’s Guide, Edition 11 9-17Executing Sample ProgramsHCI Measurement ProgramHCI Measurement Program This section explains how to use a
9-18 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsHCI Measurement ProgramTable 9-5 HCI Degradation Measurement Flow Step 2. Determi
Agilent E5250A User’s Guide, Edition 11 9-19Executing Sample ProgramsHCI Measurement ProgramStep 3. Selects devices valid for the HCI testThis step ju
9-20 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsHCI Measurement ProgramUser Function Setup: Display Setup: Also the DATA VARIAVLE
Agilent E5250A User’s Guide, Edition 11 9-21Executing Sample ProgramsHCI Measurement ProgramStep 5. Applies stress and characterizes the parametersThi
9-22 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsHCI Measurement ProgramThe measurement data is saved to the diskette in disk driv
Agilent E5250A User’s Guide, Edition 11 9-23Executing Sample ProgramsHCI Measurement ProgramAdding the Bias Source Control Routine This sample program
1-12 Agilent E5250A User’s Guide, Edition 11IntroductionOptions and AccessoriesOptions and Accessories This section lists the options and accessories
9-24 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsHCI Measurement ProgramSetting up the Measurement Environment Before executing th
Agilent E5250A User’s Guide, Edition 11 9-25Executing Sample ProgramsHCI Measurement ProgramFigure 9-3 Internal Connections and DIP SW Settings for E5
9-26 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsHCI Measurement ProgramFigure 9-4 DC HCI Degradation Test Equipment Connections N
Agilent E5250A User’s Guide, Edition 11 9-27Executing Sample ProgramsHCI Measurement ProgramExecuting the Program You should have already added the bi
9-28 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsHCI Measurement Program!-- Drain stress voltage should be determined by Id-Vd cha
Agilent E5250A User’s Guide, Edition 11 9-29Executing Sample ProgramsHCI Measurement ProgramTo Execute HCI Test Program1. Load the HCI test program (S
9-30 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsHCI Measurement ProgramAfter each interim characterization, the fractional change
Agilent E5250A User’s Guide, Edition 11 9-31Executing Sample ProgramsHCI Measurement ProgramFigure 9-7 Initial/Interim Characterization Example Figure
9-32 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsHCI Measurement ProgramTo Execute Analysis ProgramBefore running the program, do
Agilent E5250A User’s Guide, Edition 11 9-33Executing Sample ProgramsHCI Measurement Program3. Select softkey for parameter for which you want to extr
Agilent E5250A User’s Guide, Edition 11 1-13IntroductionOptions and AccessoriesTable 1-4 lists accessories furnished with the E5250A, E5252A, and E525
9-34 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsHCI Measurement ProgramNOTE If you don't want program to pause, change the p
Agilent E5250A User’s Guide, Edition 11 9-35Executing Sample ProgramsHCI Measurement ProgramModifying the Program This section explains how to modify
9-36 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsHCI Measurement ProgramTo change the 4155/4156 setup file usedTo change the 4155/
Agilent E5250A User’s Guide, Edition 11 9-37Executing Sample ProgramsHCI Measurement ProgramTo change Idlin and Gmmax extraction conditionTo change Ga
9-38 Agilent E5250A User’s Guide, Edition 11Executing Sample ProgramsHCI Measurement ProgramTo change the test limitsEach test device is judged to be
10 Specifications
10-2 Agilent E5250A User’s Guide, Edition 11SpecificationsThis chapter lists specifications, typical data and supplemental data for the Agilent E5250A
Agilent E5250A User’s Guide, Edition 11 10-3SpecificationsGeneral SpecificationsGeneral Specifications Temperature RangeOperating: 5 °C to 40 °CStorag
10-4 Agilent E5250A User’s Guide, Edition 11SpecificationsOption 001 Matrix SwitchOption 001 Matrix SwitchSwitch SpecificationsThis specification is d
Agilent E5250A User’s Guide, Edition 11 10-5SpecificationsOption 001 Matrix SwitchTypical DataOffset Current*1 < 0.1 pA (Low Leakage I-V Port) <
DECLARATION OF CONFORMITY According to EN ISO/IEC 17050-1:2004Manufacturer’s Name: Agilent Technologies Singapore (International) Pte. Ltd. Manufactur
1-14 Agilent E5250A User’s Guide, Edition 11IntroductionOptions and AccessoriesContents of the E5250A Program Disk:Virtual Front Panel Utility Intera
10-6 Agilent E5250A User’s Guide, Edition 11SpecificationsOption 501 MultiplexerOption 501 MultiplexerSwitch SpecificationsThis specification is defin
Agilent E5250A User’s Guide, Edition 11 10-7SpecificationsOption 501 MultiplexerTypical DataOffset Current*1 < 0.1 pA (Low Leakage I-V Port)Channe
10-8 Agilent E5250A User’s Guide, Edition 11SpecificationsAccessoriesAccessories The following specifications are applied for the accessories listed b
Agilent E5250A User’s Guide, Edition 11 10-9SpecificationsSupplemental Data when used with Agilent 4155/4156Supplemental Data when used with Agilent 4
10-10 Agilent E5250A User’s Guide, Edition 11SpecificationsSupplemental Data when used with Agilent 4155/4156
11 Error Messages
11-2 Agilent E5250A User’s Guide, Edition 11Error MessagesThis chapter lists and describes the error messages for Agilent E5250A. An error message con
Agilent E5250A User’s Guide, Edition 11 11-3Error MessagesStandard SCPI Error MessagesStandard SCPI Error Messages Standard SCPI error messages have n
11-4 Agilent E5250A User’s Guide, Edition 11Error MessagesStandard SCPI Error Messages-103 Invalid separator An illegal character was received when a
Agilent E5250A User’s Guide, Edition 11 11-5Error MessagesStandard SCPI Error Messages-120 Numeric data error An error was detected in a numeric param
Agilent E5250A User’s Guide, Edition 11 1-15IntroductionOptions and AccessoriesTable 1-5 lists accessories available for the E5250A.Table 1-5 Availabl
11-6 Agilent E5250A User’s Guide, Edition 11Error MessagesStandard SCPI Error Messages-148 Character data not allowed A character parameter is not all
Agilent E5250A User’s Guide, Edition 11 11-7Error MessagesStandard SCPI Error MessagesExecution Error If syntax of a SCPI command header and parameter
11-8 Agilent E5250A User’s Guide, Edition 11Error MessagesStandard SCPI Error MessagesDevice-Dependent Errors -3XX errors indicate that an E5250A oper
Agilent E5250A User’s Guide, Edition 11 11-9Error MessagesStandard SCPI Error MessagesQuery Errors If the output queue control of the E5250A detects o
11-10 Agilent E5250A User’s Guide, Edition 11Error MessagesE5250A Specific Error MessagesE5250A Specific Error Messages These are the E5250A-specific
Agilent E5250A User’s Guide, Edition 11 11-11Error MessagesE5250A Specific Error MessagesE5250A Channel Related Errors Error Number Error Message and
11-12 Agilent E5250A User’s Guide, Edition 11Error MessagesE5250A Specific Error MessagesE5250A Card/Mode/Port Related Errors Error Number Error Messa
Agilent E5250A User’s Guide, Edition 11 11-13Error MessagesE5250A Specific Error Messages3013 Cannot connect multiple channels in SROUTe mode Incorrec
11-14 Agilent E5250A User’s Guide, Edition 11Error MessagesE5250A Specific Error Messages
12 SCPI Command Summary
1-16 Agilent E5250A User’s Guide, Edition 11IntroductionOptions and Accessories
12-2 Agilent E5250A User’s Guide, Edition 11SCPI Command SummaryThis chapter is a quick reference for the SCPI subsystem commands available for Agilen
Agilent E5250A User’s Guide, Edition 11 12-3SCPI Command SummaryCommand Description :DIAG:TEST:CARD:CLE card_numberClears relay test result (pass/fail
12-4 Agilent E5250A User’s Guide, Edition 11SCPI Command Summary[:ROUT]:BIAS:CHAN:ENAB:CARD card_numberBias enables all channels in the specified card
Agilent E5250A User’s Guide, Edition 11 12-5SCPI Command Summary[:ROUT]:CLOS:CARD? card_number Returns channel_list of all closed channels for the spe
12-6 Agilent E5250A User’s Guide, Edition 11SCPI Command Summary[:ROUT]:CONN:SEQ? card_number Returns the connections sequence mode of the specified c
Agilent E5250A User’s Guide, Edition 11 12-7SCPI Command Summary[:ROUT]:COUP[:STAT]? card_number Returns Couple Mode of specified card. 0: Couple Mode
12-8 Agilent E5250A User’s Guide, Edition 11SCPI Command Summary:SYST:CDES? card_number Returns a description of the specified card: model number and
2 Installation
2-2 Agilent E5250A User’s Guide, Edition 11InstallationThis chapter describes requirements to install Agilent E5250A and the tasks for installation, a
Agilent E5250A User’s Guide, Edition 11 2-3InstallationTo Inspect the E5250A and AccessoriesTo Inspect the E5250A and Accessories When you receive and
2-4 Agilent E5250A User’s Guide, Edition 11InstallationRequirementsRequirements This section describes the following requirements for the E5250A.• “Po
Agilent E5250A User’s Guide, Edition 11 2-5InstallationRequirements• Plug: BS 1363/A, 250 V, 10 A• PN: 8120-4420• Plug: AS 3112, 250 V, 10 A• PN: 8120
2-6 Agilent E5250A User’s Guide, Edition 11InstallationRequirementsWARNING For protection from electrical shock, the power cable ground must not be de
Agilent E5250A User’s Guide, Edition 11 2-7InstallationInstalling the E5250AInstalling the E5250A This section describes how to install the E5250A.• “
• HerstellerbescheinigungGEÄUSCHEMISSIONLpa < 70 dBam Arbeitsplatznormaler Betriebnach DIN 45635 T. 19• Manufacturer’s DeclarationACOUSTIC NOISE EM
2-8 Agilent E5250A User’s Guide, Edition 11InstallationInstalling the E5250ATo Install Plug-in CardThe E5252As and E5255As are plug-in cards (modules)
Agilent E5250A User’s Guide, Edition 11 2-9InstallationInstalling the E5250AFigure 2-1 Module Extractor and Location of the Screw Hole To Install Blan
2-10 Agilent E5250A User’s Guide, Edition 11InstallationInstalling the E5250ATo Configure E5255AThe Agilent E5255A has three 2-input 8-output multiple
Agilent E5250A User’s Guide, Edition 11 2-11InstallationInstalling the E5250AFigure 2-2 Agilent E5255A Component Locations To Modify BIAS InputEach mu
2-12 Agilent E5250A User’s Guide, Edition 11InstallationInstalling the E5250AFor example, if a mainframe installs four E5255As which the IVin1 cable i
Agilent E5250A User’s Guide, Edition 11 2-13InstallationInstalling the E5250AE5255A Configuration ExamplesThe following example configurations are des
2-14 Agilent E5250A User’s Guide, Edition 11InstallationInstalling the E5250A4. Set bit 1 to ON (1) and bit 2 thru 4 to OFF (0) for DIP SW1 to SW3.5.
Agilent E5250A User’s Guide, Edition 11 2-15InstallationInstalling the E5250A5. Connect wires between BIASout1 and BIASin2, also between BIASout2 and
2-16 Agilent E5250A User’s Guide, Edition 11InstallationInstalling the E5250A96-output multiplexer (1-BIAS input, 1-IV input, 4 modules)To make this m
Agilent E5250A User’s Guide, Edition 11 2-17InstallationInstalling the E5250ATo Set GPIB Address Every device on the GPIB bus must have a unique addre
Safety SummaryThe following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failur
2-18 Agilent E5250A User’s Guide, Edition 11InstallationMaintenanceMaintenanceMaintenance should be performed periodically to keep the E5250A in good
3 Executing Self-Test and Leak Test
3-2 Agilent E5250A User’s Guide, Edition 11Executing Self-Test and Leak TestThis chapter describes how to execute the Self-Test and the Leak Test.The
Agilent E5250A User’s Guide, Edition 11 3-3Executing Self-Test and Leak TestExecuting Self-TestExecuting Self-Test The Self-Test of the E5250A consist
3-4 Agilent E5250A User’s Guide, Edition 11Executing Self-Test and Leak TestExecuting Self-TestTo Execute Self-Test (Standalone) Controller Test is ex
Agilent E5250A User’s Guide, Edition 11 3-5Executing Self-Test and Leak TestExecuting Self-TestNOTE If Fail LED is ON after Relay Test After executing
3-6 Agilent E5250A User’s Guide, Edition 11Executing Self-Test and Leak TestExecuting Self-TestTo Execute Self-Test using External Controller You can
Agilent E5250A User’s Guide, Edition 11 3-7Executing Self-Test and Leak TestExecuting Self-TestTable 3-2 SCPI Commands for Self-TestNOTE Front Panel I
3-8 Agilent E5250A User’s Guide, Edition 11Executing Self-Test and Leak TestExecuting Self-TestSelf-Test Programming ExampleThis sample program is eff
Agilent E5250A User’s Guide, Edition 11 3-9Executing Self-Test and Leak TestExecuting Self-Test430 ENTER @Hp5250;A 440 IF A=0 THEN GOTO 470 450 DISP
Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the instrument. R
3-10 Agilent E5250A User’s Guide, Edition 11Executing Self-Test and Leak TestUsing the Self-Test UtilityUsing the Self-Test Utility The Self-Test Util
Agilent E5250A User’s Guide, Edition 11 3-11Executing Self-Test and Leak TestUsing the Self-Test UtilityTo Start the Self-Test Utility 1. Before turni
3-12 Agilent E5250A User’s Guide, Edition 11Executing Self-Test and Leak TestUsing the Self-Test UtilityTo Execute Self-Test On the MAIN MENU of the E
Agilent E5250A User’s Guide, Edition 11 3-13Executing Self-Test and Leak TestUsing the Self-Test UtilityExecuting the Controller TestTo execute the co
3-14 Agilent E5250A User’s Guide, Edition 11Executing Self-Test and Leak TestUsing the Self-Test UtilityExecuting the Relay TestTo execute the Relay T
Agilent E5250A User’s Guide, Edition 11 3-15Executing Self-Test and Leak TestUsing the Self-Test UtilityTo Execute Leak Test NOTE Leak Test is for che
3-16 Agilent E5250A User’s Guide, Edition 11Executing Self-Test and Leak TestUsing the Self-Test UtilityTo execute the Leak Test, do the following:1.
Agilent E5250A User’s Guide, Edition 11 3-17Executing Self-Test and Leak TestUsing the Self-Test Utility4. Select desired IVn softkey (n=1 to 6) for i
3-18 Agilent E5250A User’s Guide, Edition 11Executing Self-Test and Leak TestUsing the Self-Test Utility5. Select the desired SMUm softkey (m=1 to 6).
Agilent E5250A User’s Guide, Edition 11 3-19Executing Self-Test and Leak TestUsing the Self-Test Utility8. Connect the cable from the 4155/4156 SMU po
Safety SymbolsThe general definitions of safety symbols used on equipment or in manuals are listed below.Instruction manual symbol: the product will b
3-20 Agilent E5250A User’s Guide, Edition 11Executing Self-Test and Leak TestUsing the Self-Test Utility
4 Setting up Measurement Environment
4-2 Agilent E5250A User’s Guide, Edition 11Setting up Measurement EnvironmentThis chapter explains how to connect your instruments to Agilent E5250A i
Agilent E5250A User’s Guide, Edition 11 4-3Setting up Measurement EnvironmentConnector PlatesConnector Plates A connector plate is the interface to co
4-4 Agilent E5250A User’s Guide, Edition 11Setting up Measurement EnvironmentConnector PlatesBlank Plate16495E Half size blank plate.This plate is us
Agilent E5250A User’s Guide, Edition 11 4-5Setting up Measurement EnvironmentConnecting 8-Channel Shielded Coaxial CableConnecting 8-Channel Shielded
4-6 Agilent E5250A User’s Guide, Edition 11Setting up Measurement EnvironmentMounting Connectors DirectlyMounting Connectors Directly Previous section
Agilent E5250A User’s Guide, Edition 11 4-7Setting up Measurement EnvironmentMounting Connectors DirectlyTable 4-1 Recommended Parts Table 4-2 Dimensi
4-8 Agilent E5250A User’s Guide, Edition 11Setting up Measurement EnvironmentMounting Connectors DirectlyTo Make Interlock Circuit The interlock circu
Agilent E5250A User’s Guide, Edition 11 4-9Setting up Measurement EnvironmentMounting Connectors DirectlyFigure 4-3 Dimensions of Interlock Switch (Ag
In This Manual This manual is a user’s guide for Agilent E5250A, and consists of the following chapters.•IntroductionProvides an overview of the E5250
4-10 Agilent E5250A User’s Guide, Edition 11Setting up Measurement EnvironmentMounting Connectors DirectlyTo Install LED CircuitThe 4155/4156 semicond
Agilent E5250A User’s Guide, Edition 11 4-11Setting up Measurement EnvironmentMounting Connectors DirectlyTo Connect Interlock Circuit to InstrumentBe
4-12 Agilent E5250A User’s Guide, Edition 11Setting up Measurement EnvironmentMounting Connectors DirectlyNOTE To Check Interlock Circuit If you use t
Agilent E5250A User’s Guide, Edition 11 4-13Setting up Measurement EnvironmentMounting Connectors DirectlyTo Connect Connectors to DUT Following figur
4-14 Agilent E5250A User’s Guide, Edition 11Setting up Measurement EnvironmentMounting Connectors DirectlyFor highly accurate current forcing and meas
Agilent E5250A User’s Guide, Edition 11 4-15Setting up Measurement EnvironmentConnecting the E5250A InputConnecting the E5250A Input This section expl
4-16 Agilent E5250A User’s Guide, Edition 11Setting up Measurement EnvironmentConnecting the E5250A InputTable 4-3 E5250A Input Connections Applicatio
Agilent E5250A User’s Guide, Edition 11 4-17Setting up Measurement EnvironmentConnecting the E5250A InputIn the table above, • AUX means CV1, CV2, HF1
4-18 Agilent E5250A User’s Guide, Edition 11Setting up Measurement EnvironmentConnecting the E5250A InputTo Connect Bias Source to the E5255A Bias Inp
Agilent E5250A User’s Guide, Edition 11 4-19Setting up Measurement EnvironmentConnecting the E5250A InputFigure 4-7 Connection of Triax-Dual Triax Ada
• SpecificationsLists specifications, typical data, supplemental data, and reference data for the E5250A, E5252A, and E5255A.•Error MessagesLists and
4-20 Agilent E5250A User’s Guide, Edition 11Setting up Measurement EnvironmentConnecting the E5250A OutputConnecting the E5250A Output This section de
Agilent E5250A User’s Guide, Edition 11 4-21Setting up Measurement EnvironmentConnecting the E5250A OutputFor Kelvin connection, use Kelvin triaxial c
4-22 Agilent E5250A User’s Guide, Edition 11Setting up Measurement EnvironmentConnecting the E5250A OutputTo Connect the E5255A Output The E5255A outp
Agilent E5250A User’s Guide, Edition 11 4-23Setting up Measurement EnvironmentMeasurement Cable LengthMeasurement Cable Length This section describes
4-24 Agilent E5250A User’s Guide, Edition 11Setting up Measurement EnvironmentMeasurement Cable LengthTable 4-6 Guard Capacitances of E5250A Measureme
5 Controlling the E5250A
5-2 Agilent E5250A User’s Guide, Edition 11Controlling the E5250AThis chapter explains how to control Agilent E5250A.• “Methods for Controlling the E5
Agilent E5250A User’s Guide, Edition 11 5-3Controlling the E5250AMethods for Controlling the E5250AMethods for Controlling the E5250A This section int
5-4 Agilent E5250A User’s Guide, Edition 11Controlling the E5250AMethods for Controlling the E5250AVFP Data Upload LibraryThis subprogram library is u
Agilent E5250A User’s Guide, Edition 11 5-5Controlling the E5250ASwitch Control BasicsSwitch Control Basics This section introduces basic information