Agilent Technologies DP235 Guia do Utilizador Página 148

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 147
148
Using the Oscilloscope
Calibrating the oscilloscope
Load the default calibration factors
Note that once the default calibration factors are loaded, all
calibrations must be done. This includes all of the calibrations in the
Self Cal menu. The calibration must be performed in the exact
sequence listed below.
NOTE: The calibration PROTECT/UNPROTECT switch on the back of logic analyzer
must be set to UNPROTECT.
1 Go to the Scope Calibration menu.
2 Select the Mode field, then select Service Cal from the pop-up
menu.
3 Select the Procedure field, then select Default Values from the
pop-up menu.
4 Select the Start field and follow the instructions on the display.
Loading the Default Calibration Factors
After you select the Start field, you can abort the calibration procedure
by selecting either the Mode or Procedure fields if the Continue field is
still displayed on the screen.
Vista de página 147
1 2 ... 143 144 145 146 147 148 149 150 151 152 153 ... 627 628

Comentários a estes Manuais

Sem comentários