Agilent Technologies DP235 Guia do Utilizador Página 106

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 105
106
Triggering Examples
Single-Machine Trigger Examples
To capture a write of known bad data to a
particular variable
The trigger specification ANDs the bad data on the data bus, the write
transaction on the status bus, and the address of the variable on the
address bus.
1 Go to the state analyzer's Trigger menu.
2 Define the terms BAD_DATA, WRITE, and VAR_ADDR to
represent the bad data value, write status, and the address of the
variable.
3 Under State Sequence Level 1, enter the following sequence
specification (use the Combination trigger term):
While storing "anystate" TRIGGER on "BAD_DATA • WRITE •
VAR_ADDR" Occurs "1" Else on "no state" go to level "1"
Capturing a Bad Write to a Variable
Vista de página 105
1 2 ... 101 102 103 104 105 106 107 108 109 110 111 ... 627 628

Comentários a estes Manuais

Sem comentários