54615B/54616B/54616C
Performance Test Record
Serial No. ______________________________________ Test by____________________________
Test Interval ____________________________________ Work Order No._____________________
Recommended Next Testing _______________________ Temperature _______________________
Calibrator Output Nominal Test Limits Result
dc 0 µV –500 µV to +500 µV _____________
5 V 4.990 V to 5.010 V _____________
delay 900 mVp-p 750 mV to 1050 mV _____________
–450 mVavg –525 mV to –375 mV _____________
54615B/16B
2.46 KHz 2.36 kHz to 2.56 kHz _____________
54616C
2.08 KHz 1.98 kHz to 2.18 kHz _____________
Voltage measurement accuracy
Range Power Supply Setting Test Limits Channel 1 Channel 2
5 V/Div 35 V 34.04 V to 35.96 V _____________ _____________
2 V/Div 14 V 13.616 V to 14.384 V _____________ _____________
1 V/Div 7 V 6.808 V to 7.192 V _____________ _____________
500 mV/Div 3.5 V 3.404 V to 3.596 V _____________ _____________
200 mV/Div 1.4 V 1.3616 V to 1.4384 V _____________ _____________
100 mV/Div 700 mV 680.8 mV to 719.2 mV _____________ _____________
50 mV/Div 350 mV 340.4 mV to 359.6 mV _____________ _____________
20 mV/Div 140 mV 136.16 mV to 143.84 mV _____________ _____________
10 mV/Div 70 mV 68.08 mV to 71.92 mV _____________ _____________
5 mV/Div 35 mV 33.66 mV to 36.34 mV _____________ _____________
2 mV/Div
14 mV 12.66 mV to 15.34 mV _____________ _____________
Bandwidth Test Limits Channel 1 Channel 2
≤ ±3 dB
_____________ _____________
Horizontal ∆t and 1/∆t accuracy
Generator Setting Test Limits Results
Frequency 10 kHz 9.96 kHz to 10.04 kHz _____________
Period 100 µs 99.59 µs to 100.41 µs _____________
Frequency 1 MHz 995.8 kHz to 1.0043 MHz _____________
Period 1 µs 995.9 ns to 1.004 µs _____________
Frequency 50 MHz 49.26 MHz to 50.76 MHz _____________
Period 20 ns 19.70 ns to 20.30 ns _____________
Frequency 500 MHz 471.67 MHz to 531.94 MHz _____________
Period
2 ns 1.880 ns to 2.120 ns _____________
Trigger sensitivity Test Limits Channel 1 Channel 2
Internal trigger 100 MHz at 0.5 divisions _____________ _____________
500 MHz at 1 division _____________ _____________
External
External trigger 500 MHz at 150 mV p-p _____________
100 MHz at 75 mV p-p _____________
Option 005 voltage measurement accuracy Test Limits Channel 1 Channel 2
Amplitude peak-to-peak
≥63.6 mVp-p _____________ _____________
4-19