Agilent Technologies 54657A Especificações Página 102

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 188
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 101
To verify bandwidth
In this test you verify bandwidth by using a power meter and power sensor to
set output of a signal generator at 1 MHz and at 500 MHz. You use the
peak-to-peak voltage at 1 MHz and at 500 MHz to verify the bandwidth
response of the oscilloscope.
Test limits:
all channels (±3 dB)
1
dc to 500 MHz
ac coupled 10 Hz to 500 MHz.
1
Upper bandwidth reduced 2MHz per degree C above 35 °C.
Table 4-5 Equipment Required
Equipment Critical specifications Recommended Model/Part
Signal generator 1 to 500 MHz at 200 mV Agilent 8656B opt 001
Power meter and
Power Sensor
1 to 500 MHz ±3% accuracy Agilent 436A and Agilent
8482A
Power splitter Outputs differ by < 0.15 dB Agilent 11667B
Cable Type N (m), 24 inch Agilent 11500B
Adapter Type N (m) to BNC (m) Agilent 1251-0082
Service
Verifying Oscilloscope Performance
4-10
Vista de página 101
1 2 ... 97 98 99 100 101 102 103 104 105 106 107 ... 187 188

Comentários a estes Manuais

Sem comentários