To display a signal automatically
The oscilloscope has an Autoscale feature that automatically sets up the
oscilloscope to best display the input signal. Using Autoscale requires signals
with a frequency greater than or equal to 50 Hz and a duty cycle greater than
0.5%.
When you press
Autoscale , the oscilloscope turns on and scales all
channels that have signals applied, and selects a time base range based
on the trigger source. The trigger source is selected from inputs that
have a signal applied. The priority of trigger source assignment is
External Trigger, input 1, then input 2. Autoscale will, in both 50 and
1M impedance modes, reset the Coupling to DC, the Bandwidth Limit
(BW Lim) to Off, all Verniers to Off, and Signal Inversion (Invert) to Off.
Input protection in 50 mode is not affected by Autoscale.
1
Connect a signal to the oscilloscope.
2 Press Autoscale .
When you press Autoscale , the oscilloscope changes the front-panel
setup to display the signal. However, if you pressed
Autoscale
unintentionally, you can use the Undo Autoscale feature. To use this
feature, perform the following step.
Press Setup , then press the Undo Autoscale softkey.
The oscilloscope returns to the configuration in effect before you
pressed
Autoscale .
The Oscilloscope at a Glance
To display a signal automatically
1–7