
7
6) PROBE COMPENSATION: This step is a “MUST DO” procedure, each and every time you
turn on an oscilloscope. In fact, you must do it if you move a probe from one channel to
another, and especially from one ‘scope to another ‘scope (for example, from the analog to
the digital ‘scope).
If you don’t make sure that each probe is compensated properly, you can wind up with
incorrect waveforms and major measurement errors. The error can be seen immediately
with pulse waveforms (like the squarewave shown below), but does NOT show up at all for
sinusoidal waveforms (AC signals). All measurements of AC signal amplitude and phase
can have big errors if compensation is not done!
In fact, performing probe compensation is so important that oscilloscope manufacturers put a
“Probe Adjust” or “Probe Comp(ensation)” output voltage on the front panel of most
oscilloscopes.
PROBE COMPENSATION PROCEDURE:
a. Turn on the oscilloscope.
b. Connect a 10X probe to the Vertical
Channel 1 BNC input connector, and
connect the probe tip to the Probe
Comp output on the lower right side
of the front panel.
c. Press the Save/Recall key, then
press the Default Setup softkey
under the display.
d. Press the Autoscale key on the front
panel. You should now see a square
wave, with 5 divisions peak-to-peak
vertically.
e. Check the waveform presentation for overshoot and rolloff (see Figure above). If necessary,
adjust the probe compensation screw on the probe assembly for flat tops on the
waveforms.
f. Disconnect the probe connected to Channel 1, and set it aside. Connect the other 10X
probe to the Vertical Channel 2 BNC input connector.
g. Press the Autoscale key on the front panel. You should now see a square wave, with 5
divisions peak-to-peak vertically.
h. Check the waveform presentation for overshoot and rolloff (see Figure above). If necessary,
adjust the probe compensation screw for flat tops on the waveforms.
Make sure that both probes, for Channel 1 and Channel 2, are properly compensated
before proceeding.
Comentários a estes Manuais