13) Another triggering method that can be used is Pattern triggering. Press the Pattern
hardkey in the Trigger Section, and use the Entry Knob to select D0. Press the L
softkey (L = logic Low), and the rotate the Entry Knob to select D1. Press the L softkey
again, and select D2; again press the L softkey. Lastly, Select D3, and press the up
arrow ( ↑↑
↑↑
) softkey, to choose positive-edge triggering on channel D3. This process will
now make the trigger occur when D2 - D0 are Low, and D3 has a positive edge. The
display should look like the one below.
14) Cursors can be used with a digital display to show the logic levels in either Binary mode
or Hex mode. Press the Cursors hardkey (in the Measure section), then press the
Mode softkey to select Binary. The X1 and X2 cursors are positioned in the display
below to show the logic values of 0111 just before the trigger point, and 1000 just after
the trigger (the time base was changed to 10 ms/div for clarity). Try Hex mode as well.