Specifications (continued)
1. This accuracy is essentially a combination of linearity, instrumentation accuracy, and traceability to absolute accuracy
at 50 MHz, 0 dBm. Note: Mismatch uncertainty, calibration factor uncertainty, and power level dependent terms (zero
set, drift, and noise) are excluded in this specification and specified elsewhere in the data sheet.
2. It is advisable to perform external zeroing on the U2000 Series power sensor for power measurement level below
–30 dBm. During the external zeroing process, the RF input signal must be switched off or the device-under-test
disconnected from the U2000 Series power sensor.
Switching point
The U2000 Series power sensors have two measurement paths: a low power path and a high power path, as shown in the table below.
Models
U2000/1/2/4A
U2000/1/2H
U2000/1B
AUTO (default) range
–60 dBm to +20 dBm
–50 dBm to +30 dBm
–30 dBm to +44 dBm
Low power path
–60 dBm to –10 dBm
–50 dBm to 0 dBm
–30 dBm to +20 dBm
High power path
–10 dBm to +20 dBm
0 dBm to +30 dBm
+20 dBm to +44 dBm
Switching point
–10 dBm
0 dBm
+20 dBm
Each power sensor automatically selects the
proper power level path. To avoid unnecessary
switching when the power level is close to
the switching point, switching point hysteresis
has been added.
Offset at switching point:
±0.5% ( ±0.02
dB) typical
Switching point hysteresis: 1 dB typical
Example with U2000 “A” suffix sensors:
Switching point for the U2000/1/2/4A sensors
is at –10 dBm. Hysteresis causes the low
power path to remain selected until
approximately –9 dBm as the power level is
increased. Above this power, the high power
path is selected. The high power path remains
selected until approximately –11 dBm is
reached as the signal level decreases. Below
this power, the low power path is selected.
Power Accuracy
Model
U2000/1/2/4A
U2000/1/2H
U2000/1B
Power range
–60 dBm to +20 dBm
–50 dBm to +30 dBm
–30 dBm to +44 dBm
Accuracy
1
(25 °C ± 10 °C)
±3.0%
±4.0%
±3.5%
Accuracy
1
(0 °C to 55 °C)
±3.5%
±5.0%
±4.0%
Specifications valid with the following conditions:
After zeroing
2
Number of averages = 1024
After 30 minutes of power-on warm-up