Agilent Technologies 8753ET Especificações Página 51

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 262
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 50
2-11
8753ET Specifications and Characteristics
Instrument Specifications
Test Port Input
Table 2-8 Test Port Input
8753ET Test Port Input
Description Specification Supplemental Information
Frequency Range
Standard 300 kHz to 3.0 GHz 10 kHz to 3 GHz, typ.
Option 006 300 kHz to 6.0 GHz 10 kHz to 6 GHz, typ.
Frequency Response (Transmission)
300 kHz to 3 GHz ±1 dB at preset power level
3 GHz to 6 GHz ±2 dB at preset power level
Impedance
Standard 50 , nominal.
Return Loss
Standard See uncorrected load match
chart.
Maximum Input Level
Transmission Port 0 dBm
Reflection Port +10 dBm
Compression See dynamic accuracy chart
Damage Level
Standard > +20 dBm or > 35 Vdc, typ.
Noise Floor
a
a. RMS value of a linear magnitude trace expressed in dBm.
8753ET Transmission Port
300 kHz to 3 GHz
3 kHz IF Bandwidth ≤ −90 dBm
10 Hz IF Bandwidth ≤ −110 dBm ≤ −120 dBm, typ.
3 GHz to 6 GHz
3 kHz IF Bandwidth ≤ −85 dBm
10 Hz IF Bandwidth ≤ −105 dBm ≤ −114 dBm, typ.