Agilent 81624B Agilent 81624B Agilent 81626B Agilent 81626B
Calibration Option C01/C02 Calibration Option C01/C02
Sensor Element InGaAs, ø 5 mm InGaAs, ø 5mm
Wavelength Range 800 nm to 1700 nm 850 nm to 1650nm
Power Range –90 dBm to +10 dBm –70 to +27 dBm (1250 nm to 1650 nm)
–70 to +23 dBm (850 nm to 1650 nm)
Applicable Fiber Type Open Beam Standard SM and MM max 100 µm core size, NA ≤0.3 Standard SM and MM max 100 µm core size, NA ≤0.3
Parallel beam max ø 4 mm Parallel beam max ø 4 mm
Uncertainty at Reference Conditions ±2.2 % ±1.5 % ±3.0 % ±2.5 %
(1000 nm to 1630 nm) (970 nm to 1630 nm) (950 nm to 1630 nm) (950 nm to 1630 nm)
Total Uncertainty ±3.5% ± 5 pW ±2.8% ± 5 pW ±5.0% ± 500 pW ±4.5% ± 500 pW
(970 nm to 1630 nm) (950 nm to 1630 nm) (950 to 1630 nm max 23 dBm) (1250 to 1630 nm max 27 dBm)
Relative Uncertainty
– due to polarization ≤ ±0.005 dB (typ. ±0.002 dB) ≤ ±0.005 dB (typ. ±0.002 dB)
– spectral ripple (due to interference) ±0.005 dB (typ. <±0.002 dB) ≤ ±0.005 dB (typ. <±0.002 dB)
Linearity (power) CW –70 dBm to +10 dBm, 1000 nm to 1630 nm CW –50 dBm to +27dBm, 950 nm to 1630 nm
– at 23°C ±5°C <±0.02 dB <±0.04 dB
– at operat. temp. range <±0.05 dB <±0.15 dB
Return Loss typ. 60 dB >45 dB >47 dB
Noise (peak to peak) <5 pW <500 pW
Averaging Time (min.) 100 µs 100 µs
Analog Output included Included
Dimensions 57 mm x 66 mm x 156 mm 57 mm x 66 mm x 156 mm
Weight 0.5 kg 0.5 kg
Recommended Recalibration Period 2 years 2 years
Operating Temperature 0°C to 40°C 0°C to +35°C
Humidity Non-condensing Non-condensing
Warm-up Time 40 min 40 min
Agilent 81628B with Integrating Sphere
Sensor Element InGaAs
Wavelength Range 800 nm to 1700 nm
Power Range –60 dBm to +40 dBm (800 nm to 1700 nm) For operation higher than 34 dBm
1
Damage Power 40.5 dBm
Applicable Fiber Type Single mode NA ≤0.2, Multimode NA ≤0.4
Open Beam ø ≤3 mm center of sphere
Uncertainty at Reference Conditions ±3.0 % (970 nm to 1630 nm)
Total Uncertainty (970 nm to 1630 nm)
≤10 dBm ±4.0% ± 5 nW
>10 dBm to ≤20 dBm ±4.5%
>20 dBm to ≤38 dBm ±5%
Relative Uncertainty
– due to polarization typ. ≤±0.006 dB
– due to speckle noise at source linewidth: typ. ≤±0.02 dB
0.1 pm to 100 pm >100 pm typ. ≤±0.002 dB
Linearity (power) (CW –40 dBm to +38 dBm), (970 nm to 1630 nm)
≤10 dBm ≤±0.03 dB
>10 dBm to ≤20 dBm ≤±0.06 dB
>20 dBm to ≤37 dBm ≤±0.09 dB
>37 dBm to ≤38 dBm ≤±0.10 dB
at 23°C ± 5°C, for operating temperature range add ±0.03 dB
Return Loss typ. >75 dB
Noise (peak to peak) <5 nW
Averaging Time (minimal) 100 µs
Analog Output Included
Dimensions 55 mm x 80 mm x 250 mm Operating Temperature 0°C to +40°C
Weight 0.9 kg (without heat sink) Humidity Non-condensing
Recommended Recalibration Period 2 years Warm-up Time 40 min
1
For optical power higher than 34 dBm the attached heat sink MUST be used! For continuous optical power or average optical power higher than 38 dBm the connector adapters will get warmer
than permitted according to the safety standard IEC 61010-1. The 81628B Optical Head can handle optical power up to 40 dBm, however, operation above 38 dBm is at the operator’s own risk.
Agilent Technologies Deutschland GmbH will not be liable for any damage caused by an operation above 38 dBm.
www.agilent.com/comms/lightwave
Response Modules
Agilent 8163xA/B and 8162xB Optical Power Meter (cont.)
22