5
Types of probes (continued)
Differential active probes
A “differential” probe is an active
probe that has two inputs, one
positive and one negative, as
well as a separate ground lead; it
drives a single-terminated 50-Ω
cable to transmit its output to one
oscilloscope channel. The output
signal is proportional to the
difference between the voltages
appearing at the two inputs. A
differential probe is used to look
at signals that are referenced
to each other instead of earth
ground and to look at small
signals in the presence of large
DC offsets or other common mode
signals such as power line noise.
Differential active probe characteristics
Model Attenuation Probe bandwidth Input dynamic range Applications and use Oscilloscope compatibility
N2772A 20:1 or 20 MHz Max input: 600 V CAT III, • For high-voltage circuits, High-impedance BNC input
200:1 1000 V CAT II motor control, power
(single-ended) or supply design
1000 Vdc, 1000 Vrms, • Requires N2773A power
1200 V DC +AC peak supply or 9-V battery
(differential)
1141A 1:1 200 MHz ±300 mV (1:1) ; • Surface-mount devices 50-Ω BNC input
±3 V (10:1) ; • Requires 1142A power
±30 V (100:1) with supply
attenuation
1153A
1
1:1 200 MHz ±300 mV (1:1) ; Surface-mount devices 50-Ω AutoProbe
±3 V (10:1) ; interface input
±30 V (100:1) with
attenuation
1 Not compatible with 5000, 6000, and 7000 Series
InfiniiMax single-ended and differential probes characteristics
Recommended
Model Attenuation Probe bandwidth Input dynamic range Applications and use oscilloscopes
1130A
1
10:1 1.5 GHz 5 V single ended, • Measure fast transitions 6000 (300 MHz-1 GHz), 7000,
±2.5 V differential 8000 Series
1131A
1
10:1 3.5 GHz 5 V single ended, DSO80204B, 80304B, 90254A
±2.5 V differential
1132A
1
10:1 5 GHz 5 V single ended, DSO80404B, 90404A
±2.5 V differential
1134A
1
10:1 7 GHz 5 V single ended, DSO80604B, 90604A
±2.5 V differential
1168A
1, 2
3.45:1 10 GHz 3.3 V single ended, DSO80804B, 81004B, 90804A
±1.65 V differential
1169A
1, 2
3.45:1 12 GHz 3.3 V single ended, DSO81204B, 81304B, 91204A
(13 GHz typical) ±1.65 V differential 91304A
1 Order one or more probe heads. See page 10 for available InfiniiMax probe heads and accessories.
2 Not compatible with 5000, 6000 and 7000 Series oscilloscopes.
• Measure fast
transitions on
low-voltage differential
or single-ended signals
• Full-bandwidth
probing system for
6000/8000/80000/90000
Series
• Requires one or more
probe head accessory
per amplifier
Comentários a estes Manuais