5. Record the DMM reading in Table 2-1 (HI to LO)
HI to Chassis Isolation 1. Make hardware connections as shown in Figure 2-7
2.
Set DMM to 2-wire ohms, 1 G range
3. Send CLOS (@10000:1rrcc) to the matrix to close all relays
where rr = highest row number and cc = highest column number
For E1465A, rr = 15 and cc = 15
For E1466A, rr = 03 and cc = 63
For E1467A, rr = 07 and cc = 31
4. Trigger the DMM with TRIG SGL
5. Record the DMM reading in Table 2-1 (HI to Chassis)
LO to Chassis Isolation 1. Make hardware connections as shown in Figure 2-8
2.
Set DMM to 2-wire ohms, 1 G range
3. Send CLOS (@10000:1rrcc) to the matrix to close all relays
where rr = highest row number and cc = highest column number
For E1465A, rr = 15 and cc = 15
For E1466A, rr = 03 and cc = 63
For E1467A, rr = 07 and cc = 31
4. Trigger the DMM with TRIG SGL
Figure 2-7. HI to Chassis Isolation Test Connections
30 Verification Tests Agilent E1465A/66A/67A Service Manual