Agilent Technologies DSOX4USBSQ USB 2.0 Manual de Serviço Página 8

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 86
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 7
8 DSOX4USBSQ USB 2.0 Signal Quality Analysis Application Electrical Testing Notes
1 Preparing for Signal Quality Analysis
Oscilloscope Requirements
The DSOX4USBSQ USB 2.0 Signal Quality Analysis application requires an
Agilent InfiniiVision 4000 X-Series oscilloscope, firmware version 3.10 or
later.
For Hi-Speed
Tests
Hi- Speed tests require a 1.5 GHz bandwidth model oscilloscope.
Hi- Speed tests on a 2- channel oscilloscope with single- ended SMA cable
connections is not supported because the sample rate will not meet the
5 GSa/s rate that is required. Using a differential probe on a 2- channel
oscilloscope works fine.
For Full and Low
Speed Tests
Full and Low Speed tests can be performed with 200 MHz, 350 MHz,
500 MHz, 1 GHz, or 1.5 GHz bandwidth model oscilloscopes.
Full Speed testing is best on a 4- channel oscilloscope because when using
a 2- channel oscilloscope, channels 1 and 2 are non- interleaved and the
sample rate will be below the 4 GSa/s that the USB-IF testing procedures
require. If Full Speed tests are run on a 2- channel oscilloscope,
interpolation is used on the data.
For Optional USB
2.0 Triggering and
Decode in the
Oscilloscope
The optional DSOX4USBH and DSOX4USBFL products (for USB 2.0
Hi- Speed and Full/Low- Speed triggering and decode, respectively) can be
very useful tools for identifying and triggering on packets that you might
want to run the USB signal quality tests on.
Vista de página 7
1 2 3 4 5 6 7 8 9 10 11 12 13 ... 85 86

Comentários a estes Manuais