7 Press the Signals softkey. In the USB Signal Quality Signals Menu:
a Press the Test Connection softkey and select whether you are using a
Single-ended (with SMA cables) or Differential probe connection.
b If you are using a differential connection, press the Source softkey
and select the analog input source channel connected to the
differential probe.
If you are using a single- ended (SMA cables) connection, press the
DP SMA softkey and select the analog input source channel connected
to the D+ signal. Then, press the DN SMA softkey and select the
analog input source channel connected to the D- signal.
On 4- channel oscilloscopes, you are forced to use different channel
pairs for the D+ and D- signals. This provides the maximum sample
rate. (Channels 1 and 2 are one pair and channels 3 and 4 are the
other pair.)
Hi- Speed tests using single- ended SMA cable connections are not
supported on 2- channel oscilloscopes because the sample rate does
not meet the required 5 GSa/s.
c Press the Test Type softkey and select Near-end or Far-end.
The terms "near-end" and "far- end" are based on which end of the
cable the test fixture is attached in relation to the device being
tested. The point where measurements are made dictates what eye
template is used.
8 Press the Auto Setup softkey to automatically set up the oscilloscope for
the selected test.
For optimal test results, you can make adjustments to the auto setup.
Generally, you want one packet of data, plus the time of one bit on
each end, across the oscilloscope's display. Also, vertical scale and
offset should be adjusted so that the signals take six vertical divisions
without being clipped.
You should see the transmitted test packet on the oscilloscope as below.