2- 18 Agilent B2900 Users Guide, Edition 2
Introduction
Specifications
Table 2-10 Voltage Measurement Specifications
Table 2-11 Current Measurement Specifications
Range
Measurement resolution
Accuracy
(gain error (% of reading)
+ offset error)
B2901A
B2902A
B2911A
B2912A
200 mV 0.1 V 0.1 V (0.015 % + 225 V)
2 V 1 V 1 V (0.02 % + 350 V)
20 V 10 V 10 V (0.015 % + 5 mV)
200 V 100 V 100 V (0.015 % + 50 mV)
Range
Measurement resolution
Accuracy
(gain error (% of reading)
+ offset error)
B2901A
B2902A
B2911A
B2912A
10 nA
a
a. 10 nA range is not available for B2901A/B2902A.
10 fA (0.10 % + 50 pA)
100 nA 100 fA 100 fA (0.06 % + 100 pA)
1 A 1 pA 1 pA (0.025 % + 500 pA)
10 A 10 pA 10 pA (0.025 % + 1.5 nA)
100 A 100 pA 100 pA (0.02 % + 25 nA)
1 mA 1 nA 1 nA (0.02 % + 200 nA)
10 mA 10 nA 10 nA (0.02 % + 2.5 V)
100 mA 100 nA 100 nA (0.02 % + 20 V)
1 A 1 A 1 A (0.03 % + 1.5 mA)
1.5 A 1 A 1 A (0.05 % + 3.5 mA)
3 A 10 A 10 A (0.4 % + 7 mA)
10 A
b
b. 10 A range is available only for pulse mode, not available for DC mode.
10 A 10 A
(0.4 % + 25 mA)
c
c. Measurement speed: 0.01 PLC