5- 28 Agilent B2900 Users Guide, Edition 2
Front Panel Operations
Performing the Limit Test
8. When the test mode is COMPLIANCE, configure the following settings:
In the Fail on field, set IN or OUT for the fail judgement method.
IN: Failure if the channel goes into the compliance state.
OUT: Failure if the channel comes out of the compliance state.
In the Fail Pattern field, set a bit pattern that represents the limit test fail state
(Fail).
9. When the test mode is LIMIT and the operation mode is GRADING, configure
the following settings:
In the Up Pattern field, set a bit pattern that represents the
failed-by-exceeding-upper-limit state.
In the Low Pattern field, set a bit pattern that represents the
failed-by-exceeding-lower-limit state.
In the Up Limit field, set the upper limit for the pass/fail judgement.
In the Low Limit field, set the lower limit for the pass/fail judgement.
10. To apply the settings, press Apply.
To apply the settings and close the dialog box, press OK.
To cancel the setting change, press the Cancel/Local key.
NOTE For using MATH in the pass/fail judgement of the limit test, see “Using the Math
Function” on page 5-25.
For using the resistance compensation, see “Enabling or Disabling the Resistance
Compensation” on page 5-24.