Agilent Technologies B2902 Guia do Utilizador Página 219

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 232
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 218
Agilent B2900 Users Guide, Edition 2 6- 31
Function Details
Trigger System
Synchronous Device Actions
This section applies only to when performing synchronous channel operation on a
2-channel model.
If channels are set as shown below, the device actions start simultaneously.
To synchronize transient actions (source output)
Trigger source is set to the same mode.
Delay time is set to the same value.
Source output ranging mode is set to the fixed mode.
Source wait time control is set to OFF.
Measurement wait time control is set to OFF.
Measurement ranging mode is set to the fixed mode.
To synchronize acquire actions (measurement)
Trigger source is set to the same mode.
Delay time is set to the same value.
Measurement wait time control is set to OFF.
Measurement ranging mode is set to the fixed mode.
Vista de página 218
1 2 ... 214 215 216 217 218 219 220 221 222 223 224 ... 231 232

Comentários a estes Manuais

Sem comentários