5- 8 Agilent B2900 User’s Guide, Edition 2
Front Panel Operations
Configuring Various Settings
Displaying the Firmware Revision
1. Press the More > System > More > Info. function keys.
2. Press Revision.
The revision information is displayed in the Revision dialog box.
3. Press OK to close the dialog box.
Setting the GPIB Address
1. Press the More > I/O function keys.
2. Press GPIB.
The GPIB Configuration dialog box is displayed.
3. Set the GPIB address.
4. To apply the settings, press Apply.
To apply the settings and close the dialog box, press OK.
To cancel the setting change, press the Cancel/Local key.
Setting the Remote Control Command Set
1. Press the More > System > More > SCPI function keys.
2. To use the default command set, press Default. To use the conventional
command set, press 2400.
3. If the command set is changed, the Confirmation dialog box is displayed. To
perform the operation, press OK. To cancel the operation, press the
Cancel/Local key.
Setting the Remote Display Mode
1. Press the More > Display function keys.
2. Press Remote.
3. To enable the screen display in remote mode, press ON. To disable it, press OFF.
To cancel the setting change, press the Cancel/Local key.