Vista de página 101
2. IF Gain
Uncertainty
Specification
Related Performance
Test
Description
0.0 dBm to -55.9 dBm . . . . . . . . . . . . . . . . . . . . . . . . . . f0.6
dB
-56.0 dBm to -129.9 dBm
. . . . . . . . . . . . . . . . . . . . . .
hl.0
dB
IF Gain Uncertainty Test
A signal source of known amplitude is connected to the spectrum
analyzer and the analyzer is adjusted for a reference level. The
amplitude of the signal peak is measured in 0.1
dB
steps from -0.1
dB
to -1.9
dB,
in 2
dB
steps from -1.9
dB
to -9.9
dB,
and in 10 steps
from -10 dB to
-
120
dB.
5-12 Operation Verification