6. On each instrument section, slide open the power module’s
fuse door (located on the rear panel). See Figure 2-6. Push the
fuse-pull lever to the left. Remove the fuse.
Note
Although Figure 2-6 shows the two instrument sections connected
together, they will be mounted separately in the rack.
Line Voltage SettingLine Voltage Setting CardCard
FuseFuse
Figure 2-6. Voltage Selector Board and Fuse
7. On each power-line module, check the voltage selector card for
the proper ac line voltage. The card’s voltage setting should
be visible on the card. See Figure 2-6. If you need to select a
different ac line voltage setting, perform the following:
a. Use needle-nose pliers to pull out the voltage-selector card.
Rotate the card so that the voltage label corresponding to the
actual line voltage will appear in the module window.
b. Push the card back into its slot.
2-10
Installation