Chapter 2 Operating Your Frequency Counter
Offsetting a Frequency Measurement
Operating Guide 2-23
2
6Press the Offset On/Off key.
The Freq Offset annunciator is activated, and the value of the display is
adjusted to reflect the value and sign of the offset entered in Steps 2 and 3.
The display should look like this:
Since the offset function can be used to add or subtract a fixed value
to/from the measurement result, you can use this feature to tune or align
the odd frequencies of a local oscillator (LO). If you enter the target
frequency as a negative offset, the Counter displays the difference
between the LO’s frequency and the target frequency. You can then adjust
the LO until the Counter displays a value of zero.
NOTE The maximum value that can be entered for Frequency Offset is
± 49,999,999,999 Hz.
GHz
MHz
kHz
Hz
Ch 1
Rel Freq
Offset