Vista de página 447
448 Chapter 16
GPIB Command Reference
SWED
SWED
Syntax SWED {UP|DOWN}
SWED?
Description Selects the sweep direction.
Parameters
Query response {UP|DOWN}<newline><^END>
Corresponding
key
[Sweep] - DIRECTION [ ]
SWET
Syntax SWET <numeric>[S]
SWET?
Description Sets the sweep time.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[Sweep] - TIME - SWEEP TIME
Description
UP (initial value) Specifies the sweep parameter increasing direction (from left to
right on the screen).
DOWN Specifies the sweep parameter decreasing direction (from right to
left on the screen).
<numeric>
Description Sweep time
Range 0 to 10000 (Note that, the lower limit varies depending on the
measurement condition.)
Initial value Varies depending on the measurement condition.
Unit s (second)
Resolution 0.001