294 Chapter 16
GPIB Command Reference
DCOMSHOR{R|L}
DCOMSHOR{R|L}
Syntax DCOMSHORR <numeric>[OHM]
DCOMSHORL <numeric>[H]
DCOMSHOR{R|L}?
Description For fixture compensation data measurement (“COM{A|B|C}” command on page 281), sets
the resistance value (“DCOMSHORR” command) and the inductance value
(“DCOMSHORL” command) of the SHORT standard used to measure SHORT
compensation data.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[Cal] - FIXTURE COMPEN - DEFINE VALUE - {SHORT RESIST(R)| SHORT INDUCT(L)}
DCRNG
Syntax DCRNG {M1|M10|M100}
DCRNG?
Description Sets the dc bias range.
Parameters
Query response {M1|M10|M100}<newline><^END>
Corresponding
key
[Source] - more 1/2 - MEAS RANGE [ ] - {1mA | 10mA | 100mA}
<numeric>
Description Resistance value or inductance value
Range -10E6 to 10E6
Initial value 0
Unit For resistance value: Ω (ohm)
For inductance value: H (henry)
Resolution 1E-12
Description
M1 (initial value) Specifies the 1-mA range.
M10 Specifies the 10-mA range.
M100 Specifies the 100-mA range.