Agilent Technologies 34450A Guia do Utilizador Página 130

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 140
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 129
108 34450A User’s Guide
4 Specifications
Supplemental measurement specifications
Tab le 4-7 Supplemental measurement specifications
DC Voltage
Measuring Method:
Sigma Delta A-to-D converter
Input Resistance:
>10 GΩ ± 2% range (Selectable 100 mV, 1 V ranges)
10 MΩ ± 2% range (typical)
Input Protection:
1000 V on all ranges (HI terminal)
Resistance
Measurement Method:
2-wire or 4-wire Ohms
Open-Circuit Voltage
Limited to < 2.8 V
Input Protection:
1000 V on all ranges (HI terminal)
DC Current
Shunt Resistance:
1 Ω for 10 mA, 100 mA
90 Ω for 100 μA, 1 mA
0.01 Ω for 1 A, 10 A
Input Protection:
Externally accessible at Rear Panel 0.4 A, 500 V fuse for I terminal
Internal 11 A, 1000 V fuse for 10 A terminal
Vista de página 129
1 2 ... 125 126 127 128 129 130 131 132 133 134 135 ... 139 140

Comentários a estes Manuais

Sem comentários