Agilent Technologies 34410A Guia do Utilizador Página 78

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 190
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 77
Description
This command selects the calculation function to be used. The default
function is NULL.
The null function provided in this command is for compatibility with
the 34401A 6½ Digit Multimeter. For the 34410A, 34411A,L4411A
it is recommended that you use the individual (null per function)
null commands found in the [SENSe:] subsystem, rather than the
null function provided by the CALCulate:NULL:OFFSet command.
Avoid using both types of nulls, as unexpected results may occur
(they are additive).
Vista de página 77
1 2 ... 73 74 75 76 77 78 79 80 81 82 83 ... 189 190