Example
The following program segment shows how to use the INITiate command
with the CONFigure and FETCh? commands. The INITiate command places
the instrument in the "wait-for-trigger" state, triggers a measurement when
the rear-panel Ext Trig Inputline is pulsed (low by default), and then sends
the readings to memory. The FETCh? command transfers the readings from
memory to the instrument's output buffer.
CONF:VOLT:DC 10,0.003
TRIG:SOUR EXT
INIT
FETC?
Typical Response: +4.27150000E-03