Remarks
This command may be useful to abort a measurement when the
instrument is waiting for a trigger, for a long measurement (for example,
100 plc), or for a long series of timed measurements.
The command will abort a measurement in progress and stop, returning
the instrument to the trigger idle state.
The *RST command will abort a measurement and set all measurement
parameters to their factory settings. The Instrument Preset
(SYSTem:PRESet command) does the same.