9
Accessory Specifications
16442A Test Fixture
Channel Information
SMU:
6 channels (1 triaxial connector/
channel)
3 channels (1 Kelvin triaxial connector/
channel)
VSU:
2 channels (1 BNC connector/channel)
VMU:
2 channels (1 BNC connector/channel)
PGU:
2 channels (1 BNC connector/channel)
GNDU:
1 channel (1 triaxial connector)
INTLK: 6 pin connector
Supplemental Information (at
23°C ± 5°C, 50% RH)
SMU channel:
Leakage current: 10pA max @200V
(Force or Sense ↔ Common)
Stray capacitance: 15pF max
(Force or Sense ↔ Common)
Stray capacitance: 3pF typical
(Force or Sense ↔ Other SMU)
Residual resistance: 60mΩ typical
(Force, Sense)
Guard capacitance: 70pF max
(Force or Sense ↔ Guard)
PGU port signal transfer charac
teristics
Overshoot: < 5% of pulse amplitude
(@20ns leading and trailing time, 50Ω
pulse generator source impedance,
50pF and 1MΩ in parallel load).
General Specifications
Dimensions:
50 mm H × 250 mm W × 275 mm D
Weight (approx.): 1.1kg
16441A R-BOX
The 16441A R-BOX adds a selectable
series resistor to the SMU output. You
can select the resistor from the setup
page, and the voltage drop due to the
series resistor is automatically
compensated for in the measurement
result.
Measurement limitations with the
4155C and 4156C and R-BOX
If you measure device characteristics
including negative resistance over
1MΩ with the 4155C/4156C and
R-BOX, there is a possibility that
they cannot be measured.
There is a possibility that the
4155C and 4156C cannot perform
measurements because of DUT
oscillations even with the R-BOX.
Whether oscillation occurs or not
depends upon the DUT and
measurement conditions.
Number of SMU channels that can add
resistor: 2
Resistor values:
1MΩ, 100kΩ, 10kΩ, 0Ω (each
channel)
Resistance accuracy:
0.3% (at 23°C±5°C, between input-
output terminal)
Maximum voltage: 200V
Maximum current: 1A (0Ω selected)
Kelvin connection: Kelvin connection
is effective only when 0Ω is selected.
Supplemental Information (at
23°C ± 5°C, 50% RH)
Leakage current: <100fA @ 100V
General Specifications
Dimensions:
72 mm H × 250 mm W × 270 mm D
Weight (approx.): 1.6kg
Specification Condition
The “supplemental information” and “typical”
entries in the following specifications are not
warranted, but provide useful information
about the functions and performance of the
instruments. 23°C ± 5°C, 50% RH.
16440A
SMU/Pulse Generator
Selector
The 16440A switches either an SMU
or PGU to the associated output port.
You can expand to 4 channels by adding
an additional 16440A. The channel 1
PGU port provides “PGU OPEN”
function, which can disconnect the PGU
by opening a semiconductor relay. The
16440A cannot work without two
pulse generator units of the
41501A/B (SMU and Pulse
Generator Expander).
Channel configurations:
Two channels (CH1, CH2)
CH1: INPUT ports: 2
(SMU and PGU, PGU port has
additional series semiconductor relay)
OUTPUT port: 1
CH2: INPUT ports: 2 (SMU and PGU)
OUTPUT port: 1
Voltage & Current Range
0.2A (AC peak)40VPGU
1.0 A200 VSMU
Max IMax. V
Input port
Supplemental Information (at
23°C ± 5°C, 50% RH)
SMU port leakage current:
< 100fA @100V
SMU port residual resistance (typical):
0.2Ω
SMU port stray capacitance (typical
@1MHz):
Force ↔ Common: 0.3pF
Force ↔ Guard: 15pF
Guard ↔ Common: 130pF
PGU port residual resistance: 3.4Ω
PGU port OFF capacitance (typical):
5pF
PGU port OPEN capacitance (typical):
700pF (@ 1MHz, Vin - Vout = 0V)
Keyboard, 1 ea.
User manual, 1 set
Sample application program disk, 1 ea.
Sample VEE program disk, 1 ea
VXIplug&play drivers disk for the
4155C and 4156C, 1 ea.
VXIplug&play drivers disk for the
E5250A, 1 ea.
4155C and 4156C
Furnished Accessories
Triaxial cable, 4 ea. (4155C)
Kelvin triaxial cable, 4 ea. (4156C)
Coaxial cable, 4 ea.
Interlock cable, 1 ea.
Altitude
Operating: 0 to 2,000 m (6,561 ft)
Storage: 0 to 4,600 m (15,091 ft)
Power requirement
90V to 264V, 47 to 63 Hz
Maximum VA
4155C and 4156C: 450VA
41501B: 350 VA
Regulatory Compliance
EMC:
EN 61326-1:+A1, AS/NZS 2064.1
Safety:
Dimensions:
4155C and 4156C:
235mm H × 426mm W × 600mm D
41501B:
190mm H × 426mm W × 600mm D
Weight (approx.):
4155C and 4156C: 21kg
41501B: 16kg
(option 412, HPSMU + 2×PGU)
CSA C22.2 NO.1010.1-1992
IEC 61010-1:+A2/EN 61010-1:+A2
UL3111-1:1994
Certification:
CE, CSA, NRTL/C, C-Tick
Comentários a estes Manuais