Agilent Technologies L4400 Guia do Utilizador Página 88

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 281
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 87
76 L4400 User’s Guide
3 Operating and Programming
You can store at least 500,000 readings in memory and all readings
are automatically time stamped. If memory overflows, the new
readings will overwrite the first (oldest) readings stored; the most
recent readings are always preserved.
You can specify a trigger count in conjunction with a sweep count.
The two parameters operate independent of one another, and the total
number of readings returned will be the product of the two
parameters.
The CONFigure and MEASure? commands automatically set the scan
trigger count to 1.
The instrument sets the scan trigger count to 1 after a Factory Reset
(
*RST command). An Instrument Preset (SYSTem:PRESet command) or
Card Reset (
SYSTem:CPON command) does not change the setting.
The following command sets the trigger count:
TRIGger:COUNt
To configure a continuous scan, send TRIG:COUNT INFINITY.
Sweep Count
Usage:
L4450A 64- Bit Digital I/O (digital input, counter channels only)
L4452A Multifunction Module (digital input, totalizer channels only)
The sweep count sets the number of sweeps per trigger event during a
scan (a sweep is one pass through the scan list).
Figure 3-11. Sweep Count Diagram.
The sweep count is valid only while scanning. If no channels have
been assigned to the scan list, the specified sweep count is ignored
(no error is generated).
t
Sweep 1Trigger
Sweep Count
(1 to 500,000 sweeps)
Sweep 2
Sweep n
Trigger
. . .
Vista de página 87
1 2 ... 83 84 85 86 87 88 89 90 91 92 93 ... 280 281

Comentários a estes Manuais

Sem comentários