L4445A Microwave Switch/Attenuator Driver 7
L4400 User’s Guide 153
SYSTem:PRESet, *RST, SYSTem:CPON and Power On
These actions drive the channels to their defined DEFault state (using the
configuration stored on the remote module) and force the system to
recognize new topologies (caused by power or connectivity changes). These
actions set the defaults shown in the table on page 152. Two parameters
are controllable to ensure safety of operation in the system; the default
s
tate for channel closure and the default state for drive enabled.
The default channel state (open or closed)
for each channel can be set
using either of the following commands.
ROUTe:CHANnel:DRIVe:CLOSe:DEFault
ROUTe:CHANnel:DRIVe:OPEN:DEFault
If a channel is configured for a single drive in pulsed mode, OPEN
operations are undefined. When these channels are configured to a
default state of
OPEN, no action is taken on these channels.
The drive state can be set as a default using the
ROUTe:RMODule:DRIVe:SOURce:BOOT command. This command allows you
to specify whether the drive current, when present, should be applied to
the switches or not. You can set
OFF, INTernal, and EXTernal for the
default.
The
*RST command forces a re- evaluation of all connected remote
modules, followed by setting all channels to their default states. This is
very similar in operation to what occurs at power- up.
ROUTe:RMODule:BANK:PRESet
This command sets a bank to default values that vary according to which
distribution board is attached. The following table shows the default states
set by
ROUTe:RMODule:BANK:PRESet.
NOTE
Y1150A Y1151A Y1152A Y1153A Y1154A Y1155A
ROUT:CHAN:DRIV:PULS:MODE ON ON ON ON ON ON
ROUT:CHAN:DRIV:PULS:WIDT 15 ms 15 ms 15 ms 15 ms 15 ms 15 ms
ROUT:CHAN:PAIR:MODE ONOFF ONONONOFF
ROUT:CHAN:DRIV:TIME:REC 0 s0 s 0 s0 s0 s0 s
ROUT:CHAN:DRIV:TIME:SETT 0 s0 s 0 s0 s0 s0 s
ROUT:CHAN:VER:ENAB OFF OFF OFF OFF OFF OFF
ROUT:CHAN:VER:POL NORM NORM INV INV NORM NORM
ROUT:RMOD:BANK:DRIV:MODE OCOL OCOL OCOL OCOL OCOL OCOL