
L4445A Microwave Switch/Attenuator Driver 7
L4400 User’s Guide 139
Figure 7- 1 shows the components of the L4445A microwave
switch/attenuator driver configuration. The L4445A driver is shown
connected to a single 34945EXT remote module.
L4445A Instrument Driver
Y1150A-Y1155A Distribution Board
9-pin D-SUB cable
(power to master 34945EXT)
34945EXT Extender
User-supplied switch
and cabling
Figure 7-1. L4445A Microwave Switch / Attenuator Driver Configuration.
• Each 3
4945EXT module can have up to four distribution boards
installed. You can have up to eight 34945EXT modules per L4445A.
• The L444
5A driver interface can supply 24 V power to the first
(master) remote module only. The first remote module can also use an
external power source..
• Sla
ve modules are connected in a daisy chain fashion using standard
ethernet RJ- 45 connectors and Cat 5 cables.
• All s
lave modules must obtain 24V power from an external power
supply. Each module can be powered by a separate supply.
• Th
e Cat 5 Ethernet cable must be plugged-in to port 1 on the master
remote module. Port 1 and Port 2 are interchangeable on all slaves.
• All distribution boar
ds on each remote module must use the same
power supply voltage.
Comentários a estes Manuais