Agilent Technologies HP8648A Manual do Utilizador Página 99

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 320
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 98
FM Receivers
98
Measurement Procedure:
1 Connect the Receiver as shown.
On the Test Set:
2 Press the PRESET key.
3 Press the RX key.
Using the knob and data entry keys:
4 Set RF Gen Freq to the receiver operating frequency.
5 Set Amplitude to 47 dBm (1 mV).
6 Set AFGen1 To 60% of the Receiver’s specified maximum frequency de-
viation (typically 3 kHz).
7 Set AFGen2 Freq to the Receiver’s CTCSS tone frequency.
8 Set AFGen2 To to the 500 Hz.
9 Set Ext Load R to the test load resistance.
10 Set AC Level meter to measure Watts.
Additional Equipment
Required
Test Load
Special Test Considerations
See "Receiver Test Loads" on page 46.
Vista de página 98
1 2 ... 94 95 96 97 98 99 100 101 102 103 104 ... 319 320

Comentários a estes Manuais

Sem comentários