Agilent Technologies HP8648A Manual do Utilizador Página 82

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 320
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 81
FM Receivers
81
Measurement Procedure:
1 Connect the Receiver as shown.
On the Test Set:
2 Press the PRESET key.
3 Press the RX key.
Using the knob and data entry keys:
4 Set RF Gen Freq to the receiver operating frequency.
5 Set Amplitude to 47 dBm (1 mV).
6 Set AFGen1 To 60% of the Receiver’s specified maximum frequency de-
viation (typically 3 kHz).
7 Set Ext Load R to the test load resistance.
8 Set AC Level meter to measure Watts.
Additional Equipment
Required
Test Load
Special Test Considerations
See "Receiver Test Loads" on page 46.