Chapter 3 65
Operation
Using an 71000 Series Spectrum Analyzer with an 70907B, 70909A, or 70910A
Operation
6. To adjust the tracking between the 71000 and the 11974, perform the following steps:
a. Set the spectrum analyzer center frequency by pressing
Freq and CENTER and
entering the corresponding start frequency value for the mixer listed below, then
save this setting in an unused register.
for an 11974A: 26.5 GHz
for an 11974Q: 33.0 GHz
for an 11974U: 40.0 GHz
for an 11974V: 50.0 GHz
b. On the rear-panel of the 11974, adjust the corresponding potentiometer for the mixer
listed below until one or both of the green LEDs are lit.
for an 11974A: 26.5 GHz SET
for an 11974Q: 33.0 GHz SET
for an 11974U: 40.0 GHz SET
for an 11974V: 50.0 GHz SET
c. Change the spectrum analyzer center frequency to the stop frequency value
indicated below, then save this setting in an unused register:
for an 11974A: 40 GHz
for an 11974Q: 50 GHz
for an 11974U: 60 GHz
for an 11974V: 75 GHz
d. On the rear-panel of the 11974, adjust the corresponding potentiometer for the mixer
listed below until one or both of the green LEDs are lit.
for an 11974A: 40 GHz SET
for an 11974Q: 50 GHz SET
for an 11974U: 60 GHz SET
for an 11974V: 75 GHz SET
e. Repeat step a through step d until green lights are lit at both frequencies without
additional adjustments.