The following procedure can be used to setup a mask template for testing the
frequency of a sine wave or a square wave. Similar methods can be used to
generate masks for testing the frequency of signals of other shapes. In the
oscilloscope setup, the vertical sensitivity and position should be adjusted so
that the amplitude is almost full scale. The trigger level should be adjusted to
the middle of the input signal. The mask template can be created by using a
function generator to generate a signal of variable frequency but of similar
shape and amplitude to the one that is going to be tested.
1 Connect the output of a function generator to the oscilloscope.
2 Set the function generator to generate a waveform with a similar
shape to the one that you are going to test.
3 Adjust the amplitude of the output until it is similar to the signal that
you are going to test.
4 Press Time on the oscilloscope, then press the Freq softkey to
turn on the automatic measurement for frequency.
5 Adjust the frequency of the output of the function generator to the
lower test limit.
The frequency can be verified by the automatic measurement.
6 Press Autostore .
7 Adjust the frequency of the output of the function generator to the
upper test limit.
An envelop of the test limit is generated.
8 Create a mask in the Define Automask menu with a tolerance of 0.0%.
For more information, refer to "To create a mask template using Autostore" in
chapter 2.
9 Specify your test region in the Mask Editor menu.
A smaller horizontal test region has a faster testing speed.
Solutions
Creating a frequency testing mask
64