Mask editing operating hint
Each limit line can have its own selectable test region. The testing time is
generally proportional to the size of the test region.
Figure 5 shows a mask that tests the overshoot of the waveform. Note that
only the part you are interested in is tested. The test region can be set
individually for the maximum and minimum limit.
Example mask template with selectable test region
Figure 5
Operating the Test Automation Module
To edit the mask to test only a portion of a waveform
35