Agilent Technologies A.18.00 Manual do Utilizador Página 283

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 581
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 282
283
Chapter 5, Advanced Operations
Status Reporting
Reading the Event Register
Syntax
STATus:HARDware2:EVENt?
Example
OUTPUT 714;"STAT:HARD2:EVEN?"
ENTER 714;Register_value
Clearing the Event Register
The EVENT register is cleared whenever it is queried or whenever the Common
Command *CLS is sent to the Test Set.
Reading the Enable Register
Syntax
STATus:HARDware2:ENABle?
Example
OUTPUT 714;"STAT:HARD2:ENAB?"
ENTER 714;Register_value
Writing the Enable Register
Syntax
STATus:HARDware2:ENABle <integer>
Example
OUTPUT 714;"STAT:HARD2:ENAB 256"
Clearing the Enable Register
The ENABLE register is cleared by writing to it with an integer value of zero.
Vista de página 282
1 2 ... 278 279 280 281 282 283 284 285 286 287 288 ... 580 581

Comentários a estes Manuais

Sem comentários