Vista de página 84
Troubleshooting 5
16800 Series Portable Logic Analyzers Service Guide 85
5 Using a digital voltmeter, measure the DC power supply
voltages. Use the following graphics when measuring the
voltages.
6 Using a voltmeter, ensure the AUX +5V power on the MIB
measures +5V. If it does not, replace the power supply.
7 Referring to the graphics that follow, note problems with
the power supply, then return to Figure 2,“System Power
Troubleshooting Flowchart,” on page 74.
Use the following graphics when measuring the voltages.