2 Displaying Data
58 Agilent 3000 Series Oscilloscopes User’s Guide
Changing the Display Settings
To display waveforms as vectors or dots
1 Press Display.
2 In the Display menu, select Typ e to toggle the waveform display
between:
Vectors — the oscilloscope connects the sample points by using digital
interpolation.
Digital interpolation maintains linearity by using a sin(x)/x digital
filter. The digital interpolation is suitable for real time sampling and
is most effective at 20 ns or faster horizontal scale settings.
Dots — the sample points are displayed.
To change the grid
1 Press Display.
2 In the Display menu, press Grid to select between:
— display grid and coordinates on the axes.
— displays coordinates on the axes.
— turns off the grid and coordinates.
Figure 26 Display Button