2 Displaying Data
42 Agilent 3000 Series Oscilloscopes User’s Guide
Horizontal position knob.
Trigger controls.
To reset the trigger offset
Turning the horizontal position knob changes the position of the trigger
point relative to the center of the screen. To return the trigger position to
the center of the screen:
1 Press Main/Delayed
.
2 In the Main/Delayed menu, press Trig-Offset Reset.
To specify a trigger holdoff
Trigger holdoff can be used to stabilize a waveform. The holdoff time is
the oscilloscope's waiting period before starting a new trigger. During the
holdoff time oscilloscope will not trigger until the holdoff has expired.
Figure 14 Trigger Holdoff