Set Up Your System 3
Agilent E6474A User’s Guide 45
Figure 16 Connecting more than one receiver to your laptop
Pulse trigger the receiver
The E645xC receivers accommodate a pulse trigger device. The
triggering device requirements are:
Cable with BNC male connector (the Agilent pulse trigger
cable input is female BNC)
Falling-edge trigger
TTL threshold
15 volt maximum pulse level
100 nanosecond minimum pulse width of period