Agilent Technologies 11970 Series Guia do Utilizador Página 102

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 101
3-42 Chapter3
Functional Testing
Scale Fidelity
Peak Search
4. Adjust the amplitude on the sythesized sweeper until the marker
amplitude on the analyzer reads 0 dBm +/– 0.10 dB. Record the
synthesized sweeper output level as the reference in Table 3-7 on
page 3-42.
NOTE 75 Input only. Adjust the synthesized sweeper amplitude until the
analyzer’s marker reads 48.8 dBmV +/– 0.10 dB.
5. On the analyzer, press the following keys:
Single
Peak Search
Marker, Delta
6. Record the marker delta reading in Table 3-7. At each new
synthesized sweeper power level, press
Single, Peak Search, and
record the marker amplitude level.
Table 3-7 Scale Fidelity
Synthesized
Sweeper Level
Minimum (dB) MarkerLevel
(dB)
Maximum (dB)
Reference=______ 0 (Reference) 0 (Reference)
Reference –4 dB –5.0 –3.0
Reference –16 dB –17.40 –15.60
Reference –28 dB –29.40 –26.60
Reference –40 dB –41.40 –38.60
Reference –52 dB –53.40 –50.60
Reference –64 dB –66.0 –62.0
Vista de página 101
1 2 ... 97 98 99 100 101 102 103 104 105 106 107 ... 320 321

Comentários a estes Manuais

Sem comentários