Timing chart and measurement time (Typical)
Timing chart of handler interface signal (Typical)
Figure 4. Timing chart of handler interface
Table 1. Value T1 through T7 (Typical)
Test Condition Timing
Screen Display R
dc
Comparator Min. Median Max.
setting meas.
T1 Trigger pulse - - - - 2µs - -
width
T2 Trigger response
time of - - - - - 0.3 ms 0.5 ms
Ready_for_Trig
T3 Trigger response
time of INDEX - - - - - 0.4 ms 0.6 ms
and EOM
T4 Measurement - - Off - - 5.7 ms 5.9 ms
time (*1) - - On - - 7.5 ms 7.7 ms
T5
Measurement data
- - - Off - 0.3 ms 0.4 ms
calculation time - - - On - 0.4 ms 0.4 ms
T6 Ready_for_Trig 1 point meas. Off - - - 0.1 ms 0.3 ms
setting time Ls-Q meas.
1 point meas. On On On - 9.8 ms 10.2 ms
Ls-Q meas.
List meas.
3 points meas. On On On - 9.8 ms 10.2 ms
Ls-Q meas.
T7 Trigger wait time - - - - 0 - -
*1: 1 point measurement, Trigger delay=0, Point delay=0
T1
T2
T3 T4 T5 T6 T7
/INDEX
/EOM
/READY_FOR_TRIG
Trigger Signal
10
Basic Measurement Characteristics