Agilent Technologies G1978B Guia do Utilizador Página 62

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 72
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 61
62 Multimode Source for 6100 Series Single Quad LC/MS User Guide
3 Reference
Guidelines for positive/negative polarity switching in mixed mode
NOTE
The above set of parameters is provided as a starting point and should yield good response
for many analytes. To obtain a specific response, i.e., high sensitivity detection of a
compound or class of compounds, or a balanced response to a wide variety of compound
classes, these parameters and the LC mobile phase composition should be examined
further using traditional techniques for optimizing a method.
Vista de página 61
1 2 ... 57 58 59 60 61 62 63 64 65 66 67 ... 71 72

Comentários a estes Manuais

Sem comentários