4-33
Replacing Instrument Assemblies
To Remove the A13 Acquisition (Option 001)
4 Slide the board to the right to free it from the locking posts.
Figure 4-50. Sliding the Board to the Right
5 Pull the board out a few inches, then disconnect the W24 ribbon cable and the cable from the SW2
Transfer Switch, as shown in Figure 4-51 and Figure 4-52.
Figure 4-51. Removing the W24 Ribbon Cable from P24 SRC_SEL
W24