2-6
Performance Verification
Time Interval Accuracy Performance Test
Figure 2-4. Setting Averaging for Measuring Time Interval Accuracy
11 On the Synthesized CW Generator, turn the RF ON/OFF to ON.
Figure 2-5. Setting Period for Measuring Time Interval Accuracy
12 On the DCA screen, press Period. See Figure 2-5.
13 On the 86100A/B, adjust the vertical scale so that the signal is on the screen.
14 On the 86100A/B, press the Clear Display hardkey.
15 Record the current DCA period readout in the Performance Test Record for each delay setting in
Table 2-1.