Agilent Technologies HP 3325B Manual de Instruções Página 55

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 54
Chapter 1 Performance Tests
Test 5: Option 030/050/124 Channel 3 Frequency Sensitivity (Agilent
53131A/132A Only)
Assembly-Level Service Guide 1-31
1
100 MHz to 2.5 GHz Sensitivity for Agilent 53131A/132A Option
030 only
1 Set the Agilent 8663B Synthesized Signal Generator to output a 2500 MHz, 40
dBm sine wave, and connect the signal to Channel 3 of the Counter as shown in
Figure 1-9.
Figure 1-9. Option 030/050/124 Channel 3 Frequency Sensitivity and Accuracy Test Setup
2 On the Counter, Press Run key.
H
P 53131A/132A
C
ounter
HP 8663A
Synthesized Signal Generator
N-to-BNC
Connector
Output
Channel 3
50
Feedthrough
Ref. In
(rear panel)
Input
(rear panel)
HP 8340B
Sweep Oscillator
Frequency Standar
d
(switch set to EXT)
Output
Vista de página 54
1 2 ... 50 51 52 53 54 55 56 57 58 59 60 ... 251 252

Comentários a estes Manuais

Sem comentários