Agilent Technologies HP 3325B Manual de Instruções Página 41

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 40
Chapter 1 Performance Tests
Test 1: Time Interval (Agilent 53131A/132A Only)
Assembly-Level Service Guide 1-17
1
Test 1: Time Interval (Agilent 53131A/132A Only)
This test verifies the Time Interval specifications between Channels 1 and 2.
Equipment
Agilent 8130A Pulse Generator (or equivalent)
Agilent 10503A BNC Cable
Counter Setup
1Cycle the POWER key to preset the Counter.
2 Press Time & Period key until TI 1 TO 2 is displayed.
3 Set CHANNEL 1 input conditions to:
Trigger/Sensitivity, AUTO TRG: OFF
Trigger/Sensitivity, LEVEL: .500V (Enter)
Trigger/Sensitivity, COMMON 1: ON
50/1M, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
4 Set CHANNEL 2 input conditions to:
Trigger/Sensitivity, AUTO TRG: OFF
Trigger/Sensitivity, LEVEL: .500V (Enter)
Trigger/Sensitivity, SLOPE: NEG
50/1M, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
Vista de página 40
1 2 ... 36 37 38 39 40 41 42 43 44 45 46 ... 251 252

Comentários a estes Manuais

Sem comentários